Saving To Memory; Measurement Range Selection - Sonel MMR-620 Operating Manual

Low-resistance meter
Hide thumbs Also See for MMR-620:
Table of Contents

Advertisement

Meas-
ure-
ments
Meas-
urement
trigger-
ing
Notes:
Continuous triggering for inductive object may be used solely when the meter is perma-
nently connected to the power grid. It is unacceptable to disconnect the meter from the
object during measurements. This is dangerous for the user and may cause damage to
the device.
- In the case of measurements of inductive objects the automatic measurement triggering
mode is not active.

3.1.4 Saving to memory

Saving
to mem-
Automatic saving of the measurement results to memory may be particularly useful in the
ory
automatic and continuous measurement triggering modes. Saving is performed directly after
the result has been displayed in the screen.
Meas-
urement

3.1.5 Measurement range selection

range
selection
For the purpose of measurement of the resistance of the object the MMR-620 meter ap-
plies measurement currents within the range between 0.1mA and10A. The measurement
range and thus the measurement current may be selected automatically or manually.
16
OPERATING MANUAL MMR-620 Revision 1.6