11
Viewing Alarm Limits
You can view the factory set alarm parameters or limits. These limits ensure that you are alerted
should your pharmaceuticals experience temperatures outside those recommended by the
Green Book, Chapter 3 page 27 and you cannot change them. Should you need them to be
01256 705 580
altered please contact our service team on
To view the parameter, access the menu by holding the arrow icons/buttons and scroll through
the options until you come to screen LПt.
H/L
P
Touch the tick icon/button to access the function and the controller will show the high/low
temperature parameters for each probe.
Reviewing the Date
To review or change the date/time enter the menu by holding the two arrow icons/buttons
simultaneously and scroll through the menu until you get to the
H/L
P
Touch the tick icon/button to access the function and use the tick icon/button to view the
year, month, day etc .If you want to change any of the settings, use the arrow icons/buttons
to change the number and hold the tick icon/button to confirm the change as described on
pages 4 and 5.
Changing the date/time may lead to data stored by the controller being overwritten.
We recommend that once the date/time is set it should not be altered unless absolutely
necessary, for example if the controller experiences a power issue and batteries aren't fitted.
If you are going to use the Micro SD card function, make sure you have downloaded the
latest data before adjusting the date or time.
or email service@labcold.com.
d-t
screen.
Min/Max Temperatures
To view your high/low temperature make the icons/buttons are visible by putting your hand
on the controller as described in page 6. Touch the high/low icon
will display the minimum and maximum temperatures recorded by probe
this process for probe to2. You can tell which probe the high and low temperatures relate to
to1
to2
because the
or
will display above the temperature displayed for the respective probe
readings.
H/L
t ° 1
P
H/L
° t 1
H/L
°
C
P
P
° t
1
t °
2
H/L
H/L
°
C
P
P
After showing the temperatures, the controller will revert to the normal display and the icons/
buttons cease to be active.
The temperatures recorded by the two probes may differ because they are taking readings
from different parts of the fridge, especially if probe
Min/Max Reset
To reset the minimum/maximum temperatures to conform with DH Green Book Guidance
page 29-30, hold the high/low
icon/button until
H/L
icon/button until the display reverts to the normal temperature screen and the minimum/
maximum temperatures will be reset. Repeat this for probe to2. However, if you don't want to
reset the temperatures, touch the mute icon/button and you will be returned to the normal
temperature display.
H/L
P
12
and the controller
H/L
to1
and repeat
° C
H/L
° t 1
H/L
°
C
P
P
° t
1
t °
2
H/L
H/L
°
C
P
P
to2
is placed in a simulated load .
CFП
flashes on the
display.
Hold the tick