13

Alarm Guide

Your IntelliCold® Pharmacy refrigerator is fitted with audio and visual alarms to help keep
the contents safe. All can be muted but unless the alarm condition is resolved, the alarm will
return after the time out interval.
The controller will show a code to explain why the fridge is alarming. An alarm code screen
alternates with the usual temperature screen. For all temperature, door and mains failure
related alarms, a red dot will illuminate, regardless of which screen is currently displaying. For
battery related alarms a battery icon will display.
t1H
- This will display when probe
the air temperature has exceeded 8oC. Because high
temperatures can adversely affect the contents of your
fridge, it should always be investigated. The alarm will also
sound when this screen is displayed.
t1L
- This will display when probe
1 l
temperature is below 2oC. Because low temperatures can
adversely affect the contents of your pharmacy fridge, it
should always be investigated. The alarm will also sound
when this screen is displayed.
t2H
- This will display when probe
2h
the temperature has exceeded 8oC. Because high
temperatures can adversely affect the contents of your
fridge, it should always be investigated. The alarm will also
sound when this screen is displayed.
t2L
- This will display when probe
temperature is below 2oC. Because low temperatures can
2l
adversely affect the contents of your pharmacy fridge, it
should always be investigated. The alarm will also sound
when this screen is displayed.
d-o
- Door open alarm which displays if the door is left
open for more than 90 seconds.The alarm will also sound.
It does not cancel when the door is shut so you need to
o
touch the mute icon/button when the screen shows d-o.
The red light will go off to show it has been cancelled.
L-b
- This warns you that the controller has detected that
the batteries are getting low. This alarm will only activate
if batteries are installed and the battery alarm is enabled.
See page 9 for instructions on how to do that. The alarm
will also sound when this screen is displayed.
n-b
- You will see this if batteries are not installed and
b
the battery alarm is enabled, such as when you first plug
the refrigerator in and haven't installed batteries (not
supplied).
t1
detects that
t1
detects that the air
The light icon/button allows you to turn the light of the refrigerator on when the light setting
(Lt) is in Auto mode.
To do so you will need to illuminate the icons/buttons by putting your hand over the controller
as described on page 6 and then touch the light icon/button. The light will turn on even if
t2
detects that
the door is shut. Please note it will not override any settings programmed by entering the
function on the settings menu so, if the light is set to
icon/button alone, you have to enter the
light
t2
detects that the
Your
unless the digital lock has been specified. To lock, push the key in gently and twist 90⁰ to the
left, to unlock twist 90⁰ to the right. The key can be safely removed from both the locked and
unlocked positions.
Keys should not be left in the lock as they protrude from the fridge. They could cause a
p
n-P
- You will see this if batteries are installed. It means
that there has been a loss of mains power. The alarm will
also sound.
If batteries are installed and running low the red dot
will display after the
to hibernating mode. This is to conserve power so
the controller can continue to record refrigerator and
temperature data for the duration of the power outage.

Light Icon/Button

Lt
menu. For information about how to change the
(Lt
function) see page 9.
H/L
t ° 1
P

Door Lock

Labcold
pharmacy and vaccine fridge is fitted as standard with a door lock with 2 keys
health and safety hazard or could be knocked, which may damage the lock.
14
n-P
alarm as the fridge goes in
Lt
on
you can't turn it off using the light
° C