3.
For current measurements up to 400 A, set the rotary switch on the adapter
clamp to the 400A position, the multimeter to the 400mV or 600mV range
and connect the banana plug of red test lead to the V jack.
4.
Press the lever to open the induction jaw.
5.
Close the jaw and place the conductor centered between the induction
clamps.
WRONG
6.
Once the reading stabilizes, read the value. If the polarity is twisted a "Minus
sign" is displayed. Use the HOLD function if the reading is difficult to read.
(Digital multimeter)

7. Maintenance

Only authorized service technicians may repair the instrument.
Note:
If the instrument is malfunctioning, please test:
- Battery condition and polarity
- Condition of the fuse(s) if available.
- Condition of the test leads.
Changing the battery(s)
Replace the battery(s) when the battery symbol or BATT is displayed on the LCD.
RIGHT
EN 7