When either of these conditions occurs, the
that the counter electrode has been disconnected from the cell.
Overload and Over Temperature conditions may both occur at the same time,
because the temperature is a consequence of power consumption, and in that case
the booster will shut down with both LEDs on.
To re-start the experiment, proceed as follows:
1. Stop the experiment from the PC software.
2. Correct the cause of the shutdown.
3. If an Over Temperature condition has occurred, wait for the booster to cool
down.
4. Restart the experiment, ensuring that the booster output remains under proper
control. The
experiment is running.
14906001_AA 12-09-07
LED will turn on to show that the cell is connected and the
CE on
Running an Experiment
LED will turn off, indicating
CE on
3-3