A Few Precautions - Sanyo MIR-153 Instruction Manual

Cooled incubator
Table of Contents

Advertisement

Downloaded from https://www.rsmanuals.com

A FEW PRECAUTIONS

O Do not use two-way sockets etc, because it
could cause problems. Also use an exclusive
breaker.
O Do not put volatile or inflammable chemicals
(like ethanol, benzine, alcohol, propane or ad-
hesive agents) in the box.
Q When cleaning, do not pour water directly on
the outside of the incubator (especially on con-
trol panel of the top surface). This could cause
a short-circuit or other problems.
0 If a completely filled bottle of material is placed
inside the cabinet and the inside temperature
is lowered to below 0°C, freezing of the contents
could break the bottle.
O When cleaning, do not use brushes, acids, ben-
zine, thinner, soap, cleaner or hot water. These
will cause discoloring or damage to coated sur-
faces. On plastic or rubber parts, they will cause
transformation, discoloration or degeneration.
Never apply volatile chemicals (like benzine etc.)
on plastic or rubber parts. When neutral deter-
gent is used, be sure to rinse thoroughly with
water afterwards.
Equip with an earth plate to protect yourself against electric shock which could result from poor insulation.
Consult our shops or electrician for these installation.
When you use this incubator in humid or wet situations, you must install a leak breaker on the power supply
side, as well as installing an earth plate. Consult our shops or an electrician for these installations.
The maximum heat load capacity acceptable to
this incubator is shown in Figures 2, 3 and 4
under the heading "TROUBLE SHOOTING." If
this limit should be exceeded. it may cause the
unit to malfunction.
How to ground earth wire
Installing a leak breaker
-4-