Introduction

This Installation Guide provides information on how to remove and replace the
Main PCB (Part Nº 804320) in a Land Instruments International Model 4500 MkIII
Continuous Opacity Monitoring System (COMS).
CAUTION
This procedure is intended for skilled instrument technicians and
should be carried out in a clean workshop.
Take appropriate precautions to prevent damage from electrostatic
discharge.
2

Removing the transceiver cover

1)
Using a 3mm Allen key, remove
the 8 screws that secure the
instrument cover.
2)
Slowly pull back the red cover
until the keyboard connector
can be seen. Disconnect the
keyboard ribbon, and then
remove the cover.