nozzle fully into handpiece (without a bur). Cover head of handpiece
with Midwest spray guard or paper towel. Spray for no more than one
(1) second. If excess debris appears, spray again for one (1) second.
2. Run handpiece (without a bur) until it reaches normal
speed, or until all excess lubricant is expelled (a minimum of
20 seconds). Expel in operatory or using Midwest Air Station. Wipe
off any excess lubricant on outside of handpiece with a dry gauze
pad or paper towel.
3. On a weekly basis
or if button gets stiff,
spray Midwest Plus
Aerosol Spray into
chuck for one second.
NOTES:
continued
• A KaVo MULTIflex lubricating nozzle will conveniently fit
onto Midwest Plus
• If a coupler is needed for use with Midwest Air Station,
order Midwest Maintenance Coupler, REF 380090.
(For 180S, 360S, 540S series.)
Cleaning Water Spray Holes
Run the cleaning wire (supplied with your handpiece) back and
forth in the water spray holes and in the air spray holes to loosen
and remove any mineral build-up at spray surface.
Aerosol Spray.
Water Holes
Air Holes
Water Hole