3. Getting Started

A. Battery Save Feature
After unpacking the LC100, it should be ready
for use. A 9 volt alkaline battery comes pre-
installed. The "Battery Save" feature is enabled
and set to 30 minutes in case the unit is acci-
dentally turned on in transit. The battery save
feature will turn off the LC100 automatically if
no keys are pressed after 30 minutes. To
change the length of time on the battery save
feature or to disable it completely perform the
following steps during power up:
1) Turn off the LC100.
2) Hold the
3) Turn on the LC100.
4) After power up, wait about 1 second and
release the
5) The display will now indicate the amount of
time the battery save feature is set for (off to
30 minutes).
6) Using the knob, set the desired time (off to
30 minutes).
7) Press
return the unit to normal mode. The stored
value will remain in memory until changed by
the operator.
B. HART™ Resistor Setup
The LC100 incorporates an internal 250 ohm
resistor to facilitate using a Rosemount 275