Service-Documentation
1.4

Disassembling the control panel (95436)

1.4.1 Open unit cover
1.4.2 Loosen marked screws
Carefully lift and turn around control head housing.
2.
SERVICING THE HOUSING COVER
2.1
Disassembling the cover
First disengage gas-loaded spring then disassemble
housing cover.
2.2
Disassembling the cover closure
Remove screws on inside of lid and lift out cover
closure.
JOBO Labortechnik GmbH + Co. KG
Kölner Str.58
D-51645 Gummersbach (Derschlag)
ATL-1500
2.3
Closure adjustment
Loosen lock nut, turn closure pin until distance to hex
nut approximates 3 thread turns. Check closing
performance.
Housing cover must press down all bottle lids.
2.4
Disassembling the cover panel
2.4.1 Disengage cover panel
3.
SERVICING THE CHEMISTRY
BOTTLES
3.1
Exchanging the bottles
Remove inner casing as explained in section 1.1. Pull
out bottle temperature sensor and remove bottle.
3.2
Exchanging the bottle lid panel
Remove inner casing according to section 1.1 of
service documentation.
- Remove screw.
- Pull fastening ring apart.
- During assembly, the fastening ring must lock in
place at bottle neck.