Servicemenue ATL 1500
1. Start of service test
• Set BCD-Switch in position 16
• Press PLUS and ENTER-Key simultaneously and the switch unit on (SET DEF
mode = setting of default values).
• Simulate a closed cover by putting a magnet to the sensor on the left side of the
control unit.
Press PLUS-Key. Display now shows:
2. Drum motor and filling pump test
Set BCD-Switch in position 1. Display shows:
• Press PLUS-Key for switching motor one step forward. Sound of working motor
should now be heard.
• Press ENTER-Key for turning on the pump - green LED "TWA" is on.
• Press ENTER-Key again for turning the pump off.
3. Lift arm test
• Set BCD-Switch in position 2. Display shows:
• Press PLUS-Key. Liftarm know has to move up, green LED is on
• Press ENTER-Key. Liftarm will move down, green LED is off.
Attention: This check is only possible, if the cover is closed or a magnet is put to the
sensor.
4. Magnet valve test
Set BCD-Switch in position 3. Display shows:
• Press ENTER-Key magnet valve will open and fill up the unit, green LED is on.
• Press ENTER-Key again magnet valve will close, green LED is off.
• Press PLUS-Key magnet valve will open and closes automatically after 10
seconds
5. Temperature Test
• Set BCD-Switch in position 5 Display shows 38ºC
• Press PLUS-Key, green Led is on heating period starts
• If temperature is reached an acoustic signal is given an de display shows END
Attention: This test can only be made if the first bottle and the water jacket is filled
with water
6. Eeprom Test
• Set BCD-Switch in position 6. Display shows:
---
FPS
Li
UA
E2P