“Warning” draws attention to application errors or actions that can lead to safety risks including the injury to persons or malfunctions – possibly even destruction of the device. Note! “Note” indicates an additional function or hint. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Never open the housing of the FTC300, especially because of the loose glass balls. When the device had been opened it may not work safely with flammable gases.
The protective earth connection must be made before all other connections. Any interruption in the protective earth can cause danger. Pay regard to the local regulations and circumstances regarding electric installations. Repairs may only be done by Messkonzept. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Cross-sensitivity means the sensitivity of the measurement on other gases than the measured component. Perturbation-sensitivity means the sensitivity of the measurement on other influences than the gas- composition, e.g. the gas pressure. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Operating Manual 2.2 Principle of Operation The FTC300 sensor determines the quantitative composition of gas mixtures by measuring the thermal conductivity between a heat source and a heat sink through the gas mixture. The measuring gas is conducted to the sensor through a stainless steel block that is kept at a constant temperature of 60°C.
96% - 100% On request Other gases and ranges on request The FTC300 must not be used with explosive gases. Flammable gases as H and CH may only be used in devices filled with glass balls. A gas mixture of a flammable gas with an inert gas in a mixing ratio such, that it is still inflammable for any amount of air added is called totally inertised.
3. Assembly of the Instrument 3.1 FTC300 Detector Unit The FTC300 detector unit consists of a hermetically sealed pressure proofed stainless steel block with a gas duct, which is suited for pressures up to 20 bar. Sample gas entering through the gas inlet is guided to the micromechanical thermal conductivity sensor and further downstream to the outlet port.
M4 cylinder head bolts are suitable. 3.3 Gas Ports On the bottom of the FTC300 housing two tubes with 6mm outer diameter for gas connection are located. They are labeled with “GAS IN” and “GAS OUT”.
The protection class of the device is only effective with all cables attached. In case cable A is not used, connector plug A has to be closed with an end fitting. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Finding the COM port the device is connected to require trial and error. The RS232 interface allows to operate the device via SetApp, see www.messkonzept.de for more information. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Multi Gas Mode allows sequential measurement of up to 16 binary gas mixtures * Dimensions: Width 145mm, Height 80mm plus connectors, Depth 85mm * Power supply 18V to 36V DC / 700mA Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
LED “System Alarm Indicator”. Setting of the conditions to trigger this relay is described in chapter 8.2.7. It is recommended to configure the “Error matrix” such that system errors of the device, see Appendix A, trigger the Status Alarm. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
The <ENTER> calls the item, that is marked as selected by the black background. Menu items are selected by the <RIGHT> key. In submenus with an “ESC”/”OK” option the <ENTER> key confirms the selection of “ESC” or “OK”. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Pressing the <UP> key during warm up switches directly to operation screen and activates the current loop. The displayed concentration value is not precise until the block is warmed up. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Selecting “Operation” as shown above and pressing <ENTER> or <UP> returns to the Operation screen. To access the submenus, select the desired submenu and press <ENTER>. The menu paths given in the following chapters start at the main menu. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
A substitute gas has at a certain concentration the same thermal conductivity as the test gas it is substituting, such it can also be used for the calibration instead. Please contact Messkonzept for details. 6.1 Set Offset Gas Concentration A1 A2 Er 0.30...
Figure 6.1 Gain Gas Menu Prior to a two point calibration the gain gas concentration has to be set in the submenu “Gain Gas”. This menu is operated analogously “Offset Gas” menu described above. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
0.00 100.00 Actu. 99.98 Calibration Actu. 0.00 ESC/OK ESC/OK ESC/OK Please Wait ESC/OK Figure 6.3 Calibrate Span Menu The menu structure is designed such that the gain calibration procedure is Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
The parameter menu gives enables to scroll through the entire parameter list. Contact Messkonzept for detailed information on the parameter. The first line contains the name of the parameter, the last line its number. To move forward in the list, press <ENTER>, to move backwards, press <Up>.
To leave the parameter menu press <RIGHT> and <ENTER>. A reset of the errors is not necessary as errors that are no longer pending are immediately removed from the error list. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
• In the “Output Setup” the properties of the external signal channels are set. • The “Expert Setup” gives write access to the parameters, reset functions, access modes and codes and test signals for subsequent connected equipment. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
1 and 4 when displaying Vol.%. The resolution in ppm is always 1ppm. All internal calculations are done in ppm. Access of measuring values via the RS-232 interface will always give the numbers in ppm with 1ppm resolution. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
The settings of relays, the scaling of the analog outputs and the measuring range do not change when using a different binary gas mixture. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
A gas flow of 80l/h leads to a gas exchange time of under 0.5s measured from the gas inlet of the device. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
<RIGHT> key. This is indicated by the active gas mixture. By pressing <ENTER> the next binary gas mixture from the “List for MGM” is selected. Pressing <RIGHT> again deactivates the gas mixture selection. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Relay 1 can be operated in different modes described below. Selecting “Relay 1” and pressing <ENTER> gives access to the different modes. 1) “R 1Mode L/H” signalling from low to high When the measured concentration increases and reaches the threshold, Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
ESC/OK Figure 8.8 Relay 1 Hysteresis Menu In the “R1 Hysteresis” menu the hysteresis value is set. The value is given in percent of the threshold value. Confirm with “OK”. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
With “Calibration frozen” the relay does not respond to any changes of the measured concentration during the calibration procedure and stays in the state it was in when the calibration began. With “Calibration active” the Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
“Relay 2” provides all functionalities of“Relay 1”, see description above. In addition Relay 2 permits the external signaling of the “Maintenance” alarm or monitoring the external auxiliary signal with the menu “Aux.”. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
If one or more of errors included in the Alarm matrix by “C. Alarm YES” are pending, the common relay triggers. In the factory settings all internal errors described in Appendix A are included in the common relay. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Adjust Loop Figure 8.13 Output Setup The FTC300 is equipped with three analog outputs. “Current Loop” is an isolated current output. “Analog Out 1” and “Analog Out 2” are not DC- isolated outputs with an output range from 0 to 10 V.
Page 38
When choosing the “Cal. active” option the currently measured value is shown at the current output during the calibration. When choosing the “Cal. frozen” the last measured value before the calibration began is shown at the current output. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
1 P134 Aux norm. Normalized signal applied to analog input 1 IOut Current Loop “AOut1-Offset” and “AOut-Gain” allow to scale the analog output according to a linear equation (Output= AOut1-Gain*signal+AOut1_Offset). Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
(e.g. with a multimeter) and enter the measured value in the third display line, then confirm with “OK” and <ENTER>. After successful adjustment the multimeter shows the value of 19.5mV. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
ESC/OK Figure 10.19 Expert Level Parameter Menu The configuration of the FTC300 is represented by a set of about 170 parameters. These parameters govern all settings and functions of the device. Press <ENTER> to scroll forward through the list and backwards by pressing <UP>.
Mode”. The operation mode is used with the setting “1.00000” in the menu “Mode Set” and “3.00000” for the safe mode. Each change in this mode requires the manual input of the actual Operator Code. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
When the device is reset to “Default Setting” the parameters relevant for the proper operation are overwritten by default values. With these values the device will not work. The correct parameters (contact Messkonzept) have to be written to the device again, see chapter 8.4.1 or the manual of SetApp under www.messkonzept.de...
11.4 V. The output “Current Loop” can be set to currents between 0 and 22mA. The analog outputs 1 and 2 can be set to supply a voltage between 0V and 10V. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
All test signals are permanently on until leaving the “I/O Test” menu. It is the responsibility of the expert to assure that the I/O test does not interfere with the subsequent connected systems and processes in an unintended way. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
SetTemp– The device still warms up below specified 0.3°C or a sudden change of range ambient temperature, gas flow disturbed the temperature control loop. Wait for a couple of minutes. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Page 47
Send the device to specified range Messkonzept with description of error. EXT. ERROR Error routed from Signal Check the external unit input “DIN” (0V=no >14V surveyed by the external error, +24V=error) error. Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
< 1% of smallest range per of flow at 80l/h 10l/h Gas pressure Max. 2000kPa (20 bar) Error due to change < 1% of smallest range per of pressure 10hPa (>800hPa) Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Page 49
Typ. Number of Switches: Power Supply: Voltage range: 18V to 36V DC Max. Current: 700mA Typical Current: 300mA Interface: RS-232 Baud rate: 19.2 kbaud Data: 8 bit Parity: None Stop: Flow control: None Messkonzept GmbH · Niedwiesenstr. 33 · 60431 Frankfurt · Germany...
Need help?
Do you have a question about the FTC300 and is the answer not in the manual?
Questions and answers