Filter should be changed every 6 months or sooner if filter discolors.
Remove filter by turning the filter cover anticlockwise and pulling out
of the unit. Discard old filter.
Replace with a new filter Additional filters should be purchased from
your provider.
Push new filter into position. Turn the filter cover clockwise to fit in
position.
You can use a small coin or flat screwdriver to open filter.
Caution: Do not reuse or wash filter or substitute any other material such
as cotton for a air-inlet filter. It will result in compressor damage.
All parts of the nebulizer kit, except tubing, should be cleaned according
to the following instructions. Your physician may specify a certain cleaning
procedure. If so, follow their recommendations.
To prevent possible risk of infection from contaminated
medication, cleaning of the nebulizer is recommended after each
aerosol treatment. Disinfecting is recommended once a day.
Clean after Every Use
With power switch in the "OFF" position, unplug power plug.
Disconnect tubing from the air-inlet connector and set aside.
Disassemble mouthpiece, or mask from cap. Open nebulizer by turning
cap counter clock wise and removing baffle.
Wash all items, except tubing, in a warm water / dishwashing detergent
solution. Rinse under warm tap water, remove detergent residue. Allow
to air dry.
Disinfect Daily
Using a clean container or bowl, soak items for 30 minutes in three parts
warm water & one part medical bacterial-germicidal disinfectant available
through your provider. With clean hands, remove items from disinfectant
solution, rinse under warm tap water, and air dry on clean paper towel.