Required Equipment - Fluke 1550C Calibration Manual

Insulation tester
Hide thumbs Also See for 1550C:
Table of Contents

Advertisement

1550C/1555
Calibration Manual

Required Equipment

Equipment required to perform the procedures in this manual is listed in
Table 2. If the recommended models are not available, equipment with
equivalent specifications may be substituted. Go to
the Users Manual for specific operating instructions.
For safe operation and maintenance of the product, have an
approved technician repair the product.
Do not attempt to use the 5500A, 5520A, or other standard
calibrator for insulation and continuity resistance tests.
Calibrator damage will result.
Equipment
HV Probe
Digital Multimeter
Load with Guard Terminal
Capacitors with Bleeder
[2]
Resistors
Calibrator
IR Cable Assembly
Calibration Software
Ammeter
Personal computer
[1]
Resistors must have a voltage coefficient consistent with the test voltage used.
Can use (3) each, 0.033 μF, 2 kV capacitors in series and (8) each, 8 μF, 450 V capacitors in series to obtain
[2]
required values. The 0.033 μF capacitors should have a 33 MΩ bleeder resistor across each capacitor. The 8 μF
capacitors should have a 15 MΩ bleeder resistor across each capacitor.
[3]
Available from
6
Table 2. Required Equipment
Minimum Required Characteristics
6 kV, ±1 % (1000:1 Divider)
11 kV, ±2 % for the 1555
500 mVdc to 1 V: ±0.02 %
Resistances
200 kΩ, ±1.25 %, 500 V
500 kΩ, ±1.25 %, 500 V
1 MΩ, ±1.25 %, 1 kV
2.5 MΩ, ±1.25 %, 2.5 kV
5 MΩ, ±1.25 %, 5 kV
[1]
10 MΩ, ±1.25 %, 10 kV
1 GΩ, ±1.25 %, 10 kV
100 GΩ, ±1.25 %, 10 kV
200 GΩ, ±1.25 %, 10 kV
500 GΩ, ±5 %, 10 kV
1 TΩ, ±5 %, 10 kV
2 TΩ, ±5 %, 10 kV
0.1 μF, ±5 %, 500 V, Polypropylene
1 μF, ±5 %, 2.5 kV, Polypropylene
DC current: 2 mA
Accuracy: ±1.25 %
DC Voltage: 0 - 550 V
Accuracy: ±0.005 %
AC Voltage: 0 - 240 V, 60 Hz
Accuracy: ±1.25 %
[3]
IBM compatible, with Microsoft Windows XP
SP2 or later + .NetFramework 2.0 or later
www.fluke.com
under 1550C/1555 Product Information.
XW Warning
W Caution
www.fluke.com
to download
Recommended
Model
Fluke 80K-6
80K-15
Fluke 8508
Combinations of:
Welwyn F Series,
Welwyn MFP2 Series
And Vishay HTS-523
Fluke 5080,
Fluke 5520A
Fluke P/N 2166275
Snorre
Fluke 8508

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1555

Table of Contents