8. The display will show the main screen and the current reading of force will be
displayed under CURRENT LOAD.
9. Verify the calibration by hanging a different calibration weight on the calibration
wire.
10. Repeat the calibration procedure if necessary.
Probe Tack Tester Operating Instructions (PT-2000)