Troubleshooting the Sprint What do you need in a kit to troubleshoot the Sprint? Batteries (2xAA) transmitting cable headset cable Signal check (part no. Z22502) lapel microphone Monitor earphones for the sprint (part no. Z43139) CL clinic contact numbers Guide to codes on LCD Check: The low battery symbol is not showing.
Page 2
Check the microphone Leave the processor on. Plug the listening headphones into the socket at the top of the processor. It is the smaller of the two sockets under the flap. You can listen to the quality of sound being picked up by the microphone (this is not the same as the signal the CI user is receiving).
Page 3
There is something wrong with the microphone. Return it to the CI centre. Or you can check if the processor is transmitting sounds to the implant by placing the coil over the back of the Sprint Turn the processor on. If it is working correctly this symbol should appear in the...
Page 4
Check the headset. Turn on the processor and set microphone sensitivity to 10 or 12. Put coil in place over implant. Speak normally into the microphone. If the segment meter moves, there is no problem with the connection between the processor and the headpiece.
Page 5
Leave it in the pouch overnight and the dry pack will absorb the excess moisture. For further information try the Cochlear website: http://www.cochlear.co.uk/support/537.asp Information on parts and accessories can be found at:...