S3. Daily Use; S3.1 Temperature Settings - Porkka ProMed MC 70 Installation And Operation Manual

Medical and laboratory refrigeration cabinets
Table of Contents

Advertisement

Available languages

Available languages

S3. Daily use

In the normal state, the display shows The internal air temperature (P1) with a +/- 2 °C
variance in normal use. The temperature is displayed with the accuracy of 1/10 degrees and
the set value is the coldest indoor air temperature. The variation goes towards warmer from
this set value.
Internal cabinet temperature displayed is affected by opening the door and changing
products, when the air temperature momentarily rises. The refrigeration is designed so that
the temperature variation will have no effect on the temperature of the stored product.
Additionally, the equipment has a separate control temperature probe (P4), which corre-
sponds with the stored product temperature.
S3.1 Temperature control (NORMAL STATE)
• Press the PROBES button and select the temperature displayed.
• P1 is indoor temperature that changes e.g. when the door is opened.
Lorem ipsum
• P4 is product probe. This is equivalent to the product temperature stored inside
the cabinet.
9
ENGLISH
Temperature
Set value