Electromagnetic Compatibility (Emc) To En60601-1-2:2007 - CareFusion BabyCO Meter Operating Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 11
Electromagnetic Compatibility (EMC) to EN60601-
1-2:2007
WARNING: use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause unexpected
or
adverse operation
The BabyCO has been tested to EN60601-1-2:2007, regarding it's ability to
operate in an environment containing other electrical/electronic equipment
(including other medical devices).
The purpose of this testing is to ensure that the BabyCO is not likely to
adversely affect the normal operation of other such equipment and that other
such equipment is not likely to adversely affect the normal operation of the
BabyCO.
Despite the testing of the BabyCO that has been undertaken, normal
operation of the BabyCO can be affected by other electrical/electronic
equipment and portable and mobile RF communications equipment.
As the BabyCO is medical equipment, special precautions are needed
regarding EMC (electromagnetic compatibility).
The BabyCo meter has an essential performance – the product should
continue to operate as intended, to an accuracy of +/- 5% of reading or
1ppm (whichever is greater)"
It is important that the BabyCO is configured and installed/put into service,
in accordance with the instructions/guidance provided herein and is used only
in the configuration as supplied.
Changes or modifications to the BabyCO may result in increased emissions or
decreased immunity of the BabyCO in relation to EMC performance.
The BabyCO should be used only with the PC serial cable provided by
Carefusion (Cat. No. 36-CAB1000, Description: Interface cable for COBRA
software). This cable should not be extended by the user. This cable should
not be used with devices other than the BabyCO. If the cable is extended by
the user, this may result in an increased level of emissions or decreased level
of immunity, in relation to the BabyCO's EMC. Use of the cable with devices
other than the BabyCO, may result in an increased level of emissions or
decreased level of immunity, in relation to the other devices's EMC.
WARNING:
The BabyCO should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use with other equipment is necessary, the
BabyCO and the other equipment should be observed / monitored, to verify
normal operation in the configuration in which it will be used.