  • ENGLISH, page 1
7.2.2 Error messages (continued)
Error displayed
Instrument reaction
Maintenance mode,
Cuvette is too hot.
cuvette heating off!
Electrode slope
Maintenance mode
Incorrect electrode
Maintenance mode
data
Filter modules are
Continued
contaminated
measurement
Drain is blocked.
Maintenance mode
Continued
Sample 1 / Sample 2
measurement
English
Cause
Solution
Check if incoming
sample is in the
specified range and
increase the cuvette
temperature setpoint
The cuvette/sample is
(CONFIGURE>
over-heated.
CUVETTE TEMP.)
Use lowest possible
cuvette temperature.
Call service if error
continue
Change membrane
and electrolyte, if there
is still a problem, check
The error occurs if the
standards,
electrode slope is not
measurement range
in the range of -50mV
and chemistry, check
to -70mV
flow rate. If all OK, but
error still present, fit
new electrode
Check cleaning
The mV value for Zero
solution (level and
related to Standard 1 is
delivery) and sample
not within the allowed
delivery, check
range.
Standard 1 (level and if
Allowed range for
correct for
(Uzero-UStandard1)
measurement range),
depending on
change membrane and
measurement range:
elektrolyte. Error may
Low (0.05 to 20 mg):
occure when sample
5mV to 200mV
has very unusual
properties. In this case
Medium (1 to 100 mg):
the error can be
20mV to 265mV
switched to "OFF" in
High (10 to 1000 mg):
the configuration
50mV to 315mV
menue.
Filter modules heavily
Clean filter modules
soiled
immediately
Check drain line. Adapt
Drain is blocked
cleaning interval to
(lime?)
water hardness.
Check sample delivery,
the amount of sample
make sure that sample
is not sufficient
line has no negative
(channel1 / channel2)
pressure, check
This occurs as error if
tightness of piston
Sample detection is set
pump, check overflow
to Error.
and air valve, check
tightness of system.
Troubleshooting
Reset error
Reset error manually
Reset error manually
Manual or
automatically when the
MEMBRANE
REPLACEMENT or
ELCTRODE
REPLACEMENT
pocess is started
Reset error manually
Reset error manually
Automatic reset when
enough sample is
available or manuell
reset
27