Download Print this page

Hach AMTAX sc Manual page 26

Addendum-operation with the sc1500 controller
Hide thumbs Also See for AMTAX sc:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Troubleshooting
7.2.2 Error messages (continued)
Error displayed
Instrument reaction
Instrument goes into
Analyzer is too cold.
the Maintenance mode
The analyzer is not
Instrument goes into
heating up.
the Maintenance mode
Maintenance mode,
Cooling failed!
starts automatically
after cooling down
Humidity in analyzer
Maintenance mode
Maintenance mode,
Humidity in sensor
the filtration probe sc is
isolated from the mains
Maintenance mode,
Filtration probe is
the filtration probe sc is
missing.
isolated from the mains
Temperature sensor is
Maintenance mode,
defective.
fan running, heating off
Cuvette sensor is
Maintenance mode,
defective.
cuvette heating off
Cuvette heating is
Continued
defective.
measurement
26
Cause
Solution
Instrument interior has
been below 4 °C
Close instrument,
(39 °F) for more than
check heating
5 minutes
Instrument cannot heat
interior adequately.
Close instrument,
(internal temperature
check heating
< 20 °C (68 °F) for
30 min)
Instrument interior is
too warm (> 47 °C
(117 °F) with electrode
temp. 45 °C (113 °F)
Check air filter and
> 52 °C (126 °F) with
clean/replace,
electrode temp. 50 °C
check fan.
(122 °F) or >57 °C
(135 °F) with electrode
temp. 55 °C (130 °F) )
There is liquid in the
Identify cause and
collecting tray
rectify
Immediately take
filtration probe sc out
of operation and
contact service. Take
There is liquid in the
the filtration probe sc
filtration probe
out of the tank and
enclosure
store the filter modules
so that they remain
moist. (See operating
instructions for the
filtration probe sc).
Immediately take
filtration probe sc out
of operation and
contact service. Take
The filtration probe sc
the filtration probe sc
is faulty or not
out of the tank and
connected
store the filter modules
so that they remain
moist. (See operating
instructions for the
filtration probe sc).
The temperature
Immediately switch off
sensor for the internal
instrument, contact
instrument temperature
service, replace main
is faulty
circuit board
The temperature
Contact service,
sensor for the cuvette
replace cuvette/sensor
is faulty
Close instrument door
The cuvette is not
and wait 10 minutes if
being heated
error accurs again,
adequately
contact service
Reset error
Reset error manually
TEST/MAINTENANCE
>RESET ERROR
Reset error manually
Manual reset or when
temperature falls 2 °C
(3.6 °F) below limit
Reset error manually
Reset error manually
Reset error manually
Reset error manually
Reset error manually
Reset error manually
English