GENOSS is an anesthesia solution injecting instrument during surgical procedures. Its injecting speed can be adjusted to help a surgery safely and accurately. Please read through this user manual before using the product to ensure safe and proper use.
Page 3
6. Function 6.1 Buttons 6.2 Announcements and Displays 6.3 Charge Display 6.4 Operation 7. Important notes 8. Cleaning and Storage 8.1 Cleaning 8.2 Storage 9. Maintenance 10. Trouble shooting 11. Label 12. Warranty...
1. Attention Special attention is marked with [ Warning!] sign. ▶ iCT injection is to be used only by doctors and specialists. ▶ Plaese read thorough this manual before using the device to ensure safe and proper use. ▶ Assemble/Disassemble the ampoule holder and the needle only when the device is OFF. ▶...
2. Package Contents English User Manual www.genoss.com www.genoss.com User Manual Main Body Anesthesia Ampoule Holder 3ea Charger Cradle Adaptor AC Cord Caution! Anesthesia cartridge holder and rechargeable battery are consumable goods.
4. Device layout ① Name Power Button ① ② Injection Button ④ ② ③ Stop Button ③ ⑧ ⑤ Display Panel ④ ⑨ Anesthesia Ampoule Holder ⑤ ⑦ Speaker ⑥ ⑦ Charge Display ⑥ DC jack ⑧ DC Plug ⑨ AC Cord ⑩...
5. Product Assembly and Disassembly 5.1 charging the Product Connect the power cord to AC adapter firmly then insert them into the DC jack on the charger cradle. ▶ Full charging takes 8 hours at the first time after purchasing this product. ▶...
Page 9
5. Product Assembly and Disassembly Main body will be inserted at the same direction as shown in the [Figure-3]. The chargedisplay will be turned on when the battery is ▶ charging, and turned off when the main body is taken out after fully charging. Refer the [Figure-5] for the color of charge display.
5. Product Assembly and Disassembly 5.2 Anesthesia Cartridge holder and a needle. Assemble of Anesthesia ampoule holder: The ampoule holder should be tightly installed at the front of the main body and turned 90 degrees to ▶ the right (clockwise) as shown in the [Figure-4]. Disassembly will be done by reverse order of installation process. [Figure-4]...
Page 11
5. Product Assembly and Disassembly Assemble of a disposable needle: The needle should be installed at the tip of the holder by turning to the clockwise. Disassembly will be done by ▶ reverse order of installation process. [Figure-5] Caution! 1. ‘Disposable needle 30G~33G’ should be used. 2.
5. Product Assembly and Disassembly 5.3 Anesthesia Cartridge Insert the injection ampoule into the ampoule holder by slotting its front as shown in the [Figure-6]. Needle will penetrate the rubber cap of the ▶ ampoule when it is inserted. [Figure-6] Press down the back of the ampoule holder until the locking system is activated to fix the ampoule holder in position as shown in the [Figure-7].
Page 13
5. Product Assembly and Disassembly Injection ampoule can be taken out when it is pushed upwards from the bottom as shown in the [Figure-8]. ▶ [Figure-8]...
6. Function 6.1 Buttons Function Operation Product Condition Repeat the turn On/Off whenever press Power Press the power button (Shortly) the Power button Remove Air Bubble Press the injection button once (Shortly) Remove air bubbles trapped in the needle 1step Press the button once (Shortly) 250sec/1.8ml (Low Speed) Change...
6. Function 6.2 Announcement and Display Indicator Situation Notification and voice guidance Indicator Situation Notification and voice guidance “Didididididing~” “Hi, you have an automatic iCT injector powered Step 3 “Beep beep beep” Power ON by electricity. Our iCT injector provides you with painless injection when you are given an Battery anesthetic”...
6. Function 6.3 Charge Display Status Explain Orange The battery is over discharged. Charging in Process The battery is charging. Charging Complete Green The battery has finished charging. In Use Colorless No luminescence [Table-5] 6.4 Operation When the power button is pressed, ‘P’ will be indicated on the panel with an announcement. ▶...
Page 17
6. Function If you press the injection button once after initialize, air bubbles in the needle will be removed. The device will then automatically be ▶ on hold. When the injection button is pressed again, the device will be activated at low speed and the button is pressed again, the speed will be adjusted to higher speed.
7. Important notes Please do not use in high temperature or humid environment. ▶ Please charge the device for more than 8 hours when it is being charged for the first time. ▶ Please use a ‘Disposable needle’ . ▶ When the device is connected to its charger with its power on, it will automatically shit off.
8. Cleaning and Storage 8.1 Cleaning Please use antibacterial solutions or ethanol solution on clean, soft wipes, to clean the device. Also, please be careful not to exert too much force ▶ or use sharp objects when cleaning. Please be careful not to contaminate the inner side of the device ▶...
9. Maintenance Check if the sheath of the adapter cable is peeled off. ▶ If problems such as unusual vibration or overheating occur, stop the ▶ device immediately and contact your service provider. ① Please take note that the device accessories and batteries contain ▶...
Page 21
9. Maintenance Check! ▶ Is there any damage to the buttons? Does the lamp on the device light up when it is being charged? Is there any contamination or damage around the charger? Is there any damage at the connection point of the Injection cartridge holder? Is there any problem with the speaker?
10. Trouble Shooting Problem Cause Solution Page Indicating lamp doesn’t turn on Press the power button The device doesn’t work. Battery is discharged Charge the device for more than 8 hours The battery has run out Change the battery The device Have a bad connection between charger Contact the provider or the manufacturer isn’t charged...
Page 23
Release the announcement at the same time Hard to press Breakdown the button by exerting Contact the provider or the manufacturer buttons much force [Table-6] ※ If the error continues after taking action as the above, Contact the GENOSS service provider.
11. Label Instructional icons: Refer to the [Table-7] ▶ Dipose terminated Type B applied part Attention/Warning electrical products Manufacturer separately [Table-7] Matter to be stated: Refer to the [Figure-10] ▶ [Figure-10]...
Page 25
11. Label Attachment location: Label is attached at the bottom of the main body such as [Figure-11] ▶ [Figure-11]...
, an anesthesia delivery system, is designed in accordance with the medical device regulation, and is made and inspected under a strict quality control system. Genoss Co., Ltd. warrants that this warranty conforms with the international warranty regulation of the consumer protection act.