6. Function
If you press the injection button once after initialize, air bubbles in the needle will be removed. The device will then automatically be
on hold. When the injection button is pressed again, the device will be activated at low speed and the button is pressed again, the speed will be
adjusted to higher speed. (ex. low speed → middle speed → high speed → low speed)
In 5 seconds after the injection is completed, the device will be initialize automatically.
The operation of the device will stop when the stop button is pressed.
After the device is stopped (with the stop button), injection button can be pressed again to resume the operation at low speed.
After the device is stopped (with the stop button) and the stop button is pressed again (for a few seconds), the device will be reset. However, if
the button is only pressed shortly, the device will not do anything with an error sound.
Caution!
Depending on the manufacturer of the anesthesia ampoule, not all the air bubbles will be removed at once. Hence, please ensure that all air bub-
bles have been removed by pressing the injection button to operate the device at low speed.