Home care bed

Advertisement

Assembly and Operation
Centre Frame Fixing:
Where the two halves of the beds meet in the middle, hand-knobs should be used to connect the
centre parts of the bed frames. The hand-knobs must be fitted through the foot-end first then into
the threaded head-end. Fully tighten the hand-knobs before commissioning the bed.
Operation:
Firstly, plug in the security key into the socket at the lower left side of the hand control.
This feature is to prevent any unauthorised use.
Controller Functions:
A - Head end up
C - Foot end up
E - Head and foot ends up
G - Bed up
IMPORTANT: Only one operation at a time is possible. The bed should not be used for more
than 2 minutes continuously in an 18 minute period. A thermal protection circuit may be
activated and the bed will not work. To reset the thermal cut out, switch off mains power
and leave for 30 minutes.
07
View from above
View from below
B - Head end down
D - Foot end down
F - Head and foot ends down
H - Bed down