Siemens ADVIA 2400 Maintenance Procedures page 36

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

CAUTION
Ensure that the level-sensor connector does not get wet to avoid damaging it.
7. Refill the bottle with cuvette conditioner.
8. Return the bottle to the same position on the shelf in the cabinet.
9. Connect the cuvette-conditioner bottle level sensor connector, then push the
connector in and turn clockwise.
10. Insert the filter and hose, then fasten the cap.
NOTE
Make sure that the filter holder is located at the bottom of the bottle.
11. Prime the fluid lines:
NOTE
If you are cleaning other bottles, wait to perform this step for all fluid lines.
a. At the Operation Panel, select
b. At the Prime Settings dialog box, select
times in all fields.
c. Select
140
.
Execute
.
Prime
and type
Prime 2
ADVIA 2400 Operator's Guide
for the Number of
10