Performing Additional Ise Electrode Washes - Siemens ADVIA 2400 Maintenance Procedures

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

6. At the WASH Set window, define the WASH2 container positions as follows:
a. Select
b. Select
c. Enter
d. Enter
e. Select

Performing additional ISE electrode washes

Materials required:
• ISE Detergent Solution
(REF 01307361, PN B01-4174-01)
Time: 5 minutes
Analyzer mode: Manual operation
ISE Detergent Solution is automatically run through the ISE module as part of the
shutdown wash procedure (WASH2). Manually perform additional ISE washes
(described in the following procedure)
conditions
• Dialysis samples are run routinely.
• The system is run more than 8 hours per day.
NOTE
Do NOT perform the ISE electrode wash more than 3 times per day (once as part of the
shutdown wash and twice on a per shift basis). Pour fresh ISE Detergent into a cup, not a
tube, before each wash, from the CTT.
1. At the Menu Panel, select
2. In the Period.wash area, select
3. At the Wash Electrode area, type the position number of the ISE Detergent container
in the Detergent posi. field.
4. In the Container field, select the type of container for the wash solution.
The recommended type of container is
5. Pour ISE Detergent Solution in the container and place it in the CTT position entered
in step 3.
6. In the Wash Electrode area, select
7. At the ISE Operation window, in the Period.wash area, select
8. Close the window, then select
Recording ISE slopes
Once a day, record the slopes from a successful ISE calibration on the Maintenance Log.
The slopes are provided on the ISE Monitor, RBL/Calibration History, and RealTime
Monitor windows following a successful calibration.
124
.
WASH2
2
for Cycles.
in the CTT cup position 1st time field and
49
2nd time field.
49
in the RTT1 and RTT2 cup positions 1st time fields and
and RTT2 cup positions 2nd time fields.
Execute
.
:
Maint
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
once per shift under either of the following
., then select
ISE Operation
OFF
, then select
Set
.
6 : 2mlCUP/Adp
Execute
, then select
when prompted.
Yes
in the CTT cup position
50
50
in the RTT1
.
.
Yes
when prompted.
ON
, then select
Set
ADVIA 2400 Operator's Guide
.