4.3.1 Scaling with Known Loads (On-Line Calibration) (Continued)
11.
Apply a known load equal to approximately 100% of the transducer range.
Press the ∂TARE button again. The meter shows the last stored value for Input 2.
12.
Press the ∂TARE button once more. The meter shows the actual signal being
13.
received.
Press the MENU button to store Input 2 value. The meter shows RD!2 (Read 2).
14.
Note
Press the ∂TARE button. The meter shows the last stored value for Read 2.
15.
Press the ßMAX button to change the value of your digits.
16.
Press the ∂TARE button to scroll horizontally to the next digit.
17.
Press the MENU button to store value as RD!2 (Read 2). The meter momentarily
18.
shows STRD , followed by RD.CF . Meter scaling is now complete.
26
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
RD!2 (Read 2) is the desired display reading at input 2.