Specifications; Battery Pack Specifications - Defibtech DBP-2009 User Manual

Automated external defibrillator (aed) rechargeable battery pack and charging stand
Table of Contents

Advertisement

9 Specifications

9.1 Battery Pack Specifications

Type
Lithium Ion (Li-ON), Rechargeable, Recyclable
11.1 VDC, 2.0 Ah, 22.2 Wh
Battery Pack Charging Time*
Less than 3 hours
Stand-by Life (installed in Defibtech AED)*
3 months
Battery Pack Useful Life
Recommended to replace Battery Pack every 3 years
(~300 charge/discharge cycles).
Capacity*
12 hours of operation in ECG monitoring mode or minimum 250 shocks
Low Battery Indicators (when installed in AED)
Visual
Audible
Environment (Operating Conditions)
Temperature: 0 to 50°C (32 to 122°F)**
Humidity: 5 to 95% (non-condensing)
Environment (Storage and Transport Conditions)
Temperature: 0 to 50°C (32 to 122°F)
Humidity: 5 to 95% (non-condensing)
Electromagnetic Compatibility
IEC 60601-1-2: 2010
The Charging Stand and the Rechargeable Battery Pack (when installed in
the AED) has the same EMC performance specifications as the Defibtech
AED powered by a non-rechargeable battery. Refer to the appropriate
Defibtech AED User Manual for details.
Physical
Weight: 3.2 ounces (90.7 g)
*Typical, fully-charged, new Battery Pack at 20°C.
**Maximum while charging: 40°C (104°F).
DAC-U2519EN-BB
10