Defibtech DDU-2000 Series Operating Manual

Defibtech DDU-2000 Series Operating Manual

Automated external defibrillator
Hide thumbs Also See for DDU-2000 Series:

Advertisement

Quick Links

Defibtech DDU-2000 Series
Automated External Defibrillator
DDU-2300
DDU-2400
DDU-2450
Operating Guide
For concise guidance on
set-up, use, maintenance
and technical specifications
ELECTRONIC
DISTRIBUTION
DAC-E2530EN-BJ

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DDU-2000 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Defibtech DDU-2000 Series

  • Page 1 Defibtech DDU-2000 Series Automated External Defibrillator • DDU-2300 • DDU-2400 • DDU-2450 Operating Guide For concise guidance on set-up, use, maintenance and technical specifications ELECTRONIC DISTRIBUTION DAC-E2530EN-BJ...
  • Page 2: Table Of Contents

    Notices Contents Defibtech, L.L.C. shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this Quick Use Instructions ..............4 material. Information in this document is subject to change without notice. Names and data When to Use (Indications, Contraindications, Summary of Primary used in any examples are fictitious unless otherwise noted.
  • Page 3: Quick Use Instructions

    User Manual Section 11.1.9 for Potential Adverse Effects of the Device on Health. OPERATOR TRAINING REQUIREMENTS In order to safely and effectively operate the DDU-2000 Series AED, a person shall have met the following requirements: • Defi btech DDU-2000 Series AED and/or defi brillation training as required by local, state, provincial, or national regulations.
  • Page 4: Diagram Of Components

    Defi btech Data Card slot Expiration Date FRONT OF AED (on back of package) Defi brillation Pads BATTERY PACK (inside of package) Defi btech DDU-2000 Series Automated External Defi brillator • DDU-2300 • DDU-2400 DEFIBRILLATION • DDU-2450 PADS PACKAGE Operating Guide...
  • Page 5: Setting Up The Aed

    SETTING UP THE AED The DDU-2000 Series AED is designed to be stored in a “ready” AED status state so that few steps are required to begin using the AED. Battery status Expires 01/20xx CHECK THE Pads status Adult Expires 06/20xx...
  • Page 6: Using The Aed

    USING THE AED If the patient is unconscious or unresponsive, and is Note : When this Information Softkey Icon is present on the screen, not breathing or not breathing normally, ensure emergency the user may press the corresponding softkey button for additional information with video instruction.
  • Page 7 USING THE AED (continued) Note : When this Information Softkey Icon is present on the screen, the user may press the corresponding softkey button for additional information with video instruction. To exit, press the softkey button again. STAND CLEAR PERFORM CPR Follow instructions to perform CPR, if needed.
  • Page 8: Using The Aed - Ecg Display (Ddu-2400 & Ddu-2450 Only)

    USING THE AED - ECG DISPLAY USING THE AED - MANUAL MODE (DDU-2400 AND DDU-2450 ONLY) (DDU-2400 ONLY) SELECTING ECG DISPLAY* The DDU-2400 AED provides a Manual Mode to override the AED features of the defi brillator. Mode of Operation The DDU-2400 and DDU-2450 allow the Manual Mode provides operator-initiated analysis, charge, shock, and disarm functions and is user to display the patient ECG when the...
  • Page 9: The Defibrillation Pads

    The connected pads package should be stored in the pad storage area on energy level (top button). the back of the DDU-2000 Series AED (see diagram on page 7). After con- necting the pads connector to the unit, push the pads package, rounded...
  • Page 10: The Battery Pack

    The optional Defi btech Data Card (DDC card) is used to store event and audio information Before inserting the battery pack into the DDU- collected by the AED. All DDU-2000 Series AEDs will operate without DDC cards and will still 2000 Series AED, ensure that the battery pack store select event information internally.
  • Page 11: Checking Aed Status

    ROUTINE MAINTENANCE ACTIVE STATUS INDICATOR (ASI) The DDU-2000 Series AED is designed to be very low maintenance. Simple maintenance tasks The Active Status Indicator (ASI) should be visually checked on a regular basis to ensure that are recommended to be performed regularly to ensure its readiness (see sample maintenance the AED is ready for use.
  • Page 12 DDU- Format data card • Do not immerse the DDU-2000 Series AED in fl uids or allow fl uids to enter the unit. 2000 Series User Manual, which can be viewed at or Run application from card •...
  • Page 13: Troubleshooting

    TROUBLESHOOTING Symptom Possible Cause Corrective Action The following table lists the symptoms, the possible causes, and the possible corrective actions for common problems. Refer to the User Manual (available at www.defi btech.com) for detailed Make sure pads connector “ Pads missing” prompt Pads not connected to unit is oriented correctly and fully explanations on how to implement the corrective actions.
  • Page 14: Warnings And Cautions

    Some very low amplitude or low frequency VT substance/air mixtures. • Do not let fl uids get into the DDU-2000 Series delivery of therapy and/or injury to the patient or rhythms may not be interpreted as shockable.
  • Page 15 • Use and store the DDU-2000 Series AED only • Follow voice prompts if the LCD screen within the range of environmental conditions becomes blank or unreadable.
  • Page 16: Technical Specifications

    IEC 61000-4-2: (Open air up to PATIENT ANALYSIS SYSTEM duration, updated specifi cation for Operating/Standby: 5%-95% 15kV or direct contact up to 8kV) DDU-2000 Series AEDs running (non-condensing) PATIENT ANALYSIS SENSITIVITY/SPECIFICITY software revision 2.4 or above. EMC (Emission)
  • Page 17: Glossary Of Symbols

    GLOSSARY OF SYMBOLS Symbol Meaning Symbol Meaning  Classified by TUV Rheinland of NA with respect to electric shock, fire, and mechanical hazard only in accordance with UL 60601-1, CAN/CSA C22.2 No.601.1-M90, IEC 60601-1, High voltage present. and IEC 60601-2-4. Conforms to UL Standard UL 60601-1. Certified to CAN/CSA Standard C22.2 No.
  • Page 18: Warranty Information

    In the event whether purchased concurrently with the defibrillator as part of replacement, Defibtech shall have the right at its sole of a configuration or separately, shall be substantially free discretion to replace the item with a new, or refurbished, same Handle with care.
  • Page 19: Contacts

    CONTACTS Manufacturer Defibtech, L.L.C. 741 Boston Post Road, Suite 201 Guilford, CT 06437 USA Y Y Y Y Y Y Y Y Tel.: 1-(866) 333-4241 (Toll-free within North America) 1-(203) 453-4507 Fax: 1-(203) 453-6657 Email: sales@defibtech.com (Sales) reporting@defibtech.com (Medical Device Reporting) service@defibtech.com...

This manual is also suitable for:

Ddu-2300Ddu-2400Ddu-2450

Table of Contents