Use an air spray for OA equipment to blow
the dust off the lens in

ILS20EP/ILS20EPUV.

Caution:
c
Do not touch the lens directly in
ILS20EP/ILS20EPUV.
G
Attach the backing to the printer.
See the SpectroProofer Mounter 24"/44"
Setup Guide
H
Set the white calibration tile to the white
calibration tile holder, and then attach it to
the printer.
See the SpectroProofer Mounter 24"/44"
Setup Guide
I
Install the mounter to the printer.
See the SpectroProofer Mounter 24"/44"
Setup Guide
J
Install ILS20EP/ILS20EPUV to the
mounter.
See the SpectroProofer Mounter 24"/44"
Setup Guide
User's Guide
Maintenance
Recertification for Color
Measurement Device
ILS20EP/ILS20EPUV
We recommend to inspect ILS20EP/ILS20EPUV
once a year.
For detailed information, contact customer
support services for assistance. If customer
support for your area is not listed below, contact
the dealer where you purchased your product.
See "Where To Get Help" on page 24
See the instructions below to send
ILS20EP/ILS20EPUV for recertification.
❏ When sending for recertification, make sure
that the white calibration tile and
ILS20EP/ILS20EPUV are in the set.
❏ Use the original box to pack the devices.
14
Caution:
c
ILS20EP/ILS20EPUV and the white
calibration tile are adjusted to be a set, and
given the same number in the factory.
When sending for recertification, make
sure to send the same numbered devices in
pairs.