Fixed Range Analysis - Teledyne 3020T Operating Instructions Manual

Trace oxygen analyzer
Hide thumbs Also See for 3020T:
Table of Contents

Advertisement

Trace Oxygen Analyzer
Use the DOWN/UP control to enter the upper value of the range (all
ranges begin at 0 ppm). Repeat for each range you want to set. ENTER to
accept the values and return to the Analysis Mode. (See note below.)
Note: The ranges must be increasing from low to high, for example,
if range 1 is set as 0–100 ppm and range 2 is set as 0–1,000
ppm, range 3 cannot be set as 0–500 ppm since it is lower than
range 2.
Ranges, alarms, and spans are always set in ppm units (over the entire
0-250,000 ppm range), even though all concentration-data outputs change
from ppm units to percent when the concentration is above 10,000 ppm.
After defining your analysis ranges, set
of the fixed ranges, as described below.

4.10.2 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the
analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, ENTER the
RANGE function by selecting RANGE from the MAIN MENU.
Use the DOWN/UP control to move
Use the DOWN/UP control to switch from
or
to set the instrument on the desired fixed range (low, medium, or
high).
ESCAPE to re-enter the ANALYZE screen using the fixed range.
NOTE:When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit, as established by
the setup, for that particular range, the output saturates at 1
V dc (or 20 mA). However, the digital readout and the RS-232
or
or
Teledyne Analytical Instruments
Operation 4
to either
or one
between the arrows.
to
,
4-17