Teledyne 3020T Operating Instructions Manual page 51

Trace oxygen analyzer
Hide thumbs Also See for 3020T:
Table of Contents

Advertisement

Trace Oxygen Analyzer
4. Are either of the alarms to be defeated?
If you are using password protection, you will need to enter your
password to access the alarm functions. Follow the instructions in section 4.4
to enter your password. Once you have clearance to proceed, ENTER the
ALARM function.
Use the DOWN/UP control to blink your choice of alarm, AL-1 or
AL-2. Then ENTER to move to the next screen.
Five parameters can be changed on this screen.
to non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to non-
alarm conditions.
The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.
The defeat function can also be used to reset a latched alarm.
(See procedures, below.)
Value of the alarm setpoint: AL–1 #### ppm (oxygen)
Out-of-range direction: HI or LO
Defeated? (Yes/No): Dft–Y/N
Failsafe? (Yes/No): Fs–Y/N
Latching? (Yes/No): Ltch–Y/N.
To define the setpoint, use ENTER to blink AL–1 ####. Then
use the DOWN/UP control to change the number. Holding the
control on the DOWN or UP position, while the number changes,
speeds up the incrementing or decrementing. (Remember, the
setpoint units are always ppm O
To set the other parameters use ENTER to blink the desired
parameter. Then use DOWN/UP to change the parameter.
Once the parameters for the alarm have been set, ENTER the
ALARM function again, and repeat this procedure for next alarm.
To reset a latched alarm, go to
two times or UP two times. (Toggle it to and then back to
Teledyne Analytical Instruments
Operation 4
.)
2
and then assert either DOWN
)
4-15

Advertisement

Table of Contents
loading

Table of Contents