Adult mask

Advertisement

Available languages

Available languages

PARI VORTEX
INSTRUCTIONS FOR USE
Single patient use (one person) only.
Product is non-sterile, BPA and latex-free.
For use only with PARI Vortex
Use with nebulizer or other product may result in inadequate treatment.
WARNING
READ ALL WARNINGS AND INSTRUCTIONS BEFORE USE. IMPROPER USE CAN
RESULT IN SERIOUS OR FATAL ILLNESS OR INJURY.
This device contains small parts that might be a choking hazard if swallowed.
Keep all small parts out of the reach of children.
PLEASE FOLLOW STEP-BY-STEP AND REFER TO ILLUSTRATIONS ON PAGE 1.
1. Setup: The PARI Vortex
mask to your Vortex
if your package did not include this adapter as seen on page 1.
a. Your mask features an expiratory valve. Follow the illustrations in Fig A and Fig B to
correctly position the valve.
b. The valve should be completely positioned on the outside, not the inside of the mask
(Fig C).
c. Firmly attach the included adapter to the mask. Firmly attach the other end of the
adapter to the mouthpiece of your Vortex
2. Taking a Treatment: Refer to your Vortex
Instructions for Use for proper Vortex
3. Cleaning: Clean mask and adapter by one of these easy ways:
a. Wash with warm water and liquid dish soap. Allow to dry, then store.
b. Place in a dishwasher basket and place on the top rack of a dishwasher. Remove and
store when cycle is complete.
4. Disinfecting: Clean as described in number 3 above.
For low-level disinfection every other treatment day, soak mask and adapter in a solution
of 1 part distilled white vinegar and 3 parts hot tap water for 1 hour. You can substitute
respiratory equipment disinfectant (like Control III
manufacturer's instructions for mixing). Rinse both parts thoroughly with warm tap water
and allow to dry before storing.
The mask and adapter can be disinfected by boiling both parts in a clean pot of water for
a full 5 minutes.
Hospitals and clinics can sterilize both parts by autoclaving (a maximum of 277° F or
136° C), follow autoclave manufacturer's instructions.
Dry both parts on a clean, lint-free cloth, then store.
EN
ADULT MASK
®
Non-Electrostatic Holding Chamber.
®
Adult Mask includes an adapter that is necessary to attach the
®
Non-Electrostatic Holding Chamber. Please contact PARI immediately
®
®
Non-Electrostatic Holding Chamber (Fig D).
®
Non-Electrostatic Holding Chamber
®
setup and operation.
) for distilled white vinegar (follow
®
Control III
is a registered trademark of Maril Products, Inc.
®
Questions? 1.800.FAST.NEB (327.8632) | PARI.com