Mastering Instructions for the LMI 200-S or LMI 200-SB to the LMI 585 Plus
1
Form: CA ???
August 01, 2005
2
R:\Quality\Calibration Instructions\CA 097 200s or sb to LMI 585 Plus.doc
Ph
(810) 714-5811
Fax (810) 714-5711
CustomerService@lmicorporation.com
3
4
Required equipment:
1.
LMI 585 Plus digital transmitter
2.
LMI 200-S or 200-SB probe
3.
LMI 6009 4 pin to 4 pin cable
4.
LMI 210 master block
Page 1 of 4

Advertisement

loading

Summary of Contents for LMI 585 Plus

  • Page 1 Fenton, MI 48430 Fax (810) 714-5711 CustomerService@lmicorporation.com Research, Development and Manufacturing of Precision Measuring Systems Mastering Instructions for the LMI 200-S or LMI 200-SB to the LMI 585 Plus Required equipment: LMI 585 Plus digital transmitter LMI 200-S or 200-SB probe...
  • Page 2 (810) 714-5811 *NOTE* These instructions are copied from the LMI 585 Plus User Guide and are intended to be used as a quick reference for the mastering process. Please refer to the LMI 585 Plus User Guide for the setup and configuration of the LMI gage.
  • Page 3 Master” step to verify zero or master. This calibration procedure will produce a negative reading when the Probe is extended beyond the 31.0 Master Point. Form: CA 068 August 01, 2005 R:\Quality\Calibration Instructions\CA 097 200s or sb to LMI 585 Plus.doc Page 3 of 4...
  • Page 4 CALSPAN to 22.00mm. The 210 Master Block will also be replaced with the 214 Master Block with a 54.0mm Master. Form: CA 068 August 01, 2005 R:\Quality\Calibration Instructions\CA 097 200s or sb to LMI 585 Plus.doc Page 4 of 4...

This manual is also suitable for:

200-sb200-s