keys move the blinking curser to the 'Operation' mode or the
Using the
or
'Utilities' mode Press
1.6.2 Language Selection
UTILITIES
EN
FR
DE
<
>
Using the
or
keys move the blinking curser to the language of choice. The
menu commands will then appear in the selected language. Press
1.6.3 Set-Up Menu
Return to Main Menu to set operation parameters.
MENU
OPERATION
<
>
keys move the blinking curser to the 'Operation' mode. Press
Using the
or
1.6.4 Setting Operating Temperature
TARGET TEMPERATURE?
25 TO 180 XXX.X C
Enter the required 'Target Temperature'. Target Temperature is the required
operating temperature. To change the Target Temperature, use the numeric
keypad and over write with the desired value.
Press
Caution: The maximum digestion temperature for polypropylene
tubes is 130 C. If higher digestion temperatures are required, use
CAUTION
the appropriate glass or Teflon replacement tubes. A full listing of
available tubes is located on Page 17.
KeyPad service manual
Rev.1.1
VER X.X
IT
ES
UTILITIES
.
12
Need help?
Do you have a question about the DigiPREP Series and is the answer not in the manual?
Questions and answers