USE PROPER PRECAUTIONS WHEN HANDLING AND SERVICING.
5.1
DISASSEMBLY
Disconnect GSI 17 from power.
Module).
b.
Disconnect and set aside all accessories.
C .
Remove 2 round control knobs, Item 1.
d.
Remove 3 slide switch caps, Item 2.
e.
Close top cover, Item 3.
f.
Turn instrument over onto its side.
Open top cover.
Remove 2 screws, Item 5, from bottom of instrument.
h.
Set instrument down.
i.
Remove 2 screws, Item 6, at back of front panel.
j.
k.
Lift front panel and printed circuit board, Items 7 and 8,
away from bottom case, Item 4.
1.
Remove screw, Item 9, in back of printed circuit board.
5.2
PARTS NUMBERS
ITEM
NUMBER
DESCRIPTION
HL & Frequency Knobs
1
Slide Switch Caps (Power, L/R,
2
Tone Type)
Top Cover (blank/no labels)
3
4
Bottom Case (blank/no labels)
#6 Screws
5
#6 Screws
6
Front Panel (blank/no labels)
7
Printed Circuit Board
8
#6 Screws
9
Front Panel Label
10
11
Frequency Dial
Handswitch
12
Power Module
13
14
Back Pack
15
Headset Assembly, TDH-39P, 50 ohms
CAUTION:
STATIC SENSITIVE DEVICES -
INSTRUCTIONS
(Battery Pack and/or Power
PART NUMBER
1717-7006 & (5220-0101)
1717-7007
1717-7002
1717-7001
7044-0037
7044-0037
1717-7000
1717-4700
7044-0037
1717-0401
1717-7003 & (1717-0402)
7874-0156
(see Power Module Sec.)
(see Battery Opt. Sec.)
8000-0175