Download Print this page

Agilent Technologies Genomic DNA ScreenTape System Quick Manual

Advertisement

Quick Links

System Components
The Agilent 2200 TapeStation system is an automated platform for simpler, faster and more reliable
electrophoresis.
It is made up of three elements:
• 2200 TapeStation System (G2964AA) or 2200 TapeStation Nucleic Acid System (G2965AA)
• Genomic DNA ScreenTape (5067- 5365) with Genomic DNA Reagents (Ladder and Sample Buffer)
(5067- 5366)
• Agilent 2200 TapeStation Software
Kit
The Genomic DNA ScreenTape assay is designed for analyzing genomic DNA samples in the size range
from 200 bp to >60000 bp.
Specifications
Analytical Specification
Sizing Range
Sensitivity
1
Sizing Precision
1
Sizing Accuracy
2
Quantitative Precision
2
Quantitative Accuracy
Linear Concentration Range
DIN functional range
Physical Specification
Analysis Time
Samples per consumable
Sample Volume Required
Shelf Life
Box/Kit size
1
Determined using the Genomic DNA Ladder as sample
2
Average result from various genomic DNA sample types
Agilent Genomic DNA ScreenTape
System Quick Guide
Genomic DNA ScreenTape Assay and Reagents
200 bp to > 60000 bp
0.5 ng/µL
200 – 15000 bp: 15 % CV
200 – 15000 bp: ±15 %
15 % CV
±20 %
10 – 100 ng/µL
5 – 300 ng/µL
16 samples: < 25 min, 96 samples: < 150 min
16
1 µL
4 months
112 samples/box
Agilent Technologies

Advertisement

loading

Summary of Contents for Agilent Technologies Genomic DNA ScreenTape System

  • Page 1 16 samples: < 25 min, 96 samples: < 150 min Samples per consumable Sample Volume Required 1 µL Shelf Life 4 months Box/Kit size 112 samples/box Determined using the Genomic DNA Ladder as sample Average result from various genomic DNA sample types Agilent Technologies...
  • Page 2 Agilent Genomic DNA ScreenTape System Quick Guide Storage Conditions • Reagent vials: 2 – 8 °C • The ScreenTape device: 2 – 8 °C (if you run less than 16 lanes, store used ScreenTape device upright at 2 – 8 °C for a maximum of 2 weeks.) •...
  • Page 3 Agilent Genomic DNA ScreenTape System Quick Guide General Information on Working with Genomic DNA • For best results ensure that all reagents are allowed to equilibrate to room temperature N O T E for 30 min prior to use. • When pipetting Sample Buffer, ensure that excess buffer droplets are removed from the tip before transfer to the sample tubes.
  • Page 4 SYBR® is a registered trademark of Molecular Probes, Inc. Technical Support For technical support, please visit www.agilent.com/genomics/contact Further Information Visit Agilent Technologies` web site. It offers useful information, support and current developments about the products and technology: www.agilent.com/genomics/tapestation *G2964-90040* Part Number: G2964-90040 Rev. D ©...