Electrical Connection

Electrical Information
The power specifications are located on the
unit identifi cation nameplate This nameplate is
permanently attached to the unit and must be
located to verify power requirements.
ivNow Power Requirements
120/220/230 V A C — 50/60 Hz, 1 ph
100W per pod (maximum of 400W
with 4 connected pods)
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
Applicable only on units with an equipotential-bonding
terminal:
To prevent an electrical shock hazard between the appliance and
other appliances or metal parts in close vicinity, an equipotential-
bonding terminal is provided. An equalization bonding lead must
be connected to this stud and the other appliances / metal parts
to provide sufficient protection against potential difference. The
terminal is marked with the following symbol.
Grounding reliability can only be achieved when
equipment is connected to an equivalent receptacle
marked "Hospital Grade."
8
NEMA 5-15P
15A - 125V Plug
Hospital Grade