Transportation and Damage Claims Environmental Conditions Transport and storage environmental conditions (not to exceed 15 days) Ambient temperature range of -40°C to +70°C (-40°F to +159°F) Relative humidity range of 10% to 95%, non-condensing Atmospheric pressure range of 50kPa to 106kPa Operational environmental conditions The appliance must acclimate to the room temperature in the environment it will be placed—24 hours is recommended The recommended environmental temperature range is 15°C to 32°C (60°F to 90°F)
Should damage occur in shipment, do not put the warmer into service until the damage has been inspected by an authorized service provider Promptly contact Pedigo customer service to report damage. 2. Read all instructions in this manual carefully before initiating the installation of this appliance.
Safety Procedures Knowledge of proper procedues is essential to the • Fluid warmers are only intended for warming medical safe operation of electrically energized appliances The solutions for irrigation and injection prior to use Refer following hazard signal words and symbols may be used to the labeling of the manufacturer of the products to throughout this manual.
Installation How to Install/Mount the ivNow ivNow modular units can be installed in a variety of ways. Up to 3 units can be linked together with a jumper cord and connected via a mounting plate. Configurations can be placed on counter top, mounted to wall or pole. The rear bracket provided with the ivNow can be used to mount the pod(s) to the wall.
Electrical Connection ivNow Power Requirements 120/220/230 V A C — 50/60 Hz, 1 ph NEMA 5-15P 15A - 125V Plug 100W per pod (maximum of 400W Hospital Grade with 4 connected pods) Safety Class I Equipment No Applied Parts Mode of Operation: Continuous Applicable only on units with an equipotential-bonding terminal: To prevent an electrical shock hazard between the appliance and...
Operation General Information Safety Feature The ivNow fluid warmer quickly warms and maintains the • The control of the ivNow is designed to display an error temperature of injection/intravenous solutions prior to their message (E-31) and stop heating if the temperature of use.
Page 10
Operation status button power status indicator display light Operating Instructions 1. Before operating the module(s), clean the exterior of the unit with a damp cloth and general hospital cleaner (isopropyl alcohol). 2. Plug the appliance into an appropriate hospital-grade receptacle. See the appliance data tag for voltage requirements.
Page 11
Operation status button power status indicator display light ivNow Status Indicators ivNow Adjustment Display the firmware version Fluid bag is not present Press and hold the POWER button for 10 seconds. The DISPLAY will show 4 dashes. Press the POWER button to return to operation. Fluid bag detected, measureing initial bag Temperature unit selection temperature...
7. Check that all display LEDs light up. Test by turning the unit off and then on All LEDs illuminate for one second at start-up. 8. Contact the Pedigo Service department for immediate repair if any problems exist.
Troubleshooting If your unit is not operating properly, check the following before calling your authorized service agent. Check the power applied to the unit. Is the plug in the outlet? Check to make sure the power is on by pressing the power button on the front of the unit.
Page 14