V-Flow Line operating instructions
Variable area flowmeters
Handbuch PCU1000
and high-precision control valves
for gases and liquids
Vögtlin Instruments AG – flow technology
Langenhagstrasse 1 | 4147 Aesch (Switzerland)
Phone 41 (0)61 756 63 00 | Fax +41 (0)61 756 63 01
www.voegtlin.com | info@voegtlin.com

Advertisement

Table of Contents
loading

Summary of Contents for Vogtlin V-Flow Series

  • Page 1 V-Flow Line operating instructions Variable area flowmeters Handbuch PCU1000 and high-precision control valves for gases and liquids Vögtlin Instruments AG – flow technology Langenhagstrasse 1 | 4147 Aesch (Switzerland) Phone 41 (0)61 756 63 00 | Fax +41 (0)61 756 63 01 www.voegtlin.com | info@voegtlin.com...
  • Page 2 V-Flow Line Operating instructions V-Flow Line Variable area flowmeters Q-Flow Variable area flowmeters V-100 High-precision control valves M-Flow Version: vflow_E4_7 For the latest information on our products, see our website at www.voegtlin.com © 2010 Vögtlin Instruments AG, Switzerland Manual Version Page V-Flow Line vflow_E4_7...
  • Page 3: Table Of Contents

    V-Flow Line Contents Introduction Service and quality Warranty General instructions Safety information Using the manual General information Operating principle of the variable area flowmeter Variable area flow meter type marking High-precision control valve type marking Technical data Technical data Q-Flow Materials Q-Flow Setups Q-Flow Technical data V-100...
  • Page 4 V-Flow Line Disassembly and maintenance Important instructions Disassembly variable area flowmeter Q-Flow Disassembly variable area flowmeter V-100 Disassembly Digiturn M-Flow Maintenance Soiling Cleaning Return Appendix Dimensions Q-Flow Dimensions V-100 Dimensions M-Flow Type code Q-Flow Type code V-100 Type code M-Flow Contamination declaration Manual Version...
  • Page 5: Introduction

    V-Flow Line Introduction We are glad that you have decided to use our variable area flow meters and high-precision control valves. Our instruments will provide you with high-quality long-lived products. This manual contains important information for commissioning and/or designing equipment. Please contact your distribution partner if anything is not clear.
  • Page 6: General Instructions

    V-Flow Line General instructions Check the package for external damage and contact your distribution partner if the instruments have visible defects. Check that the delivery is complete and corresponds to the delivery note. This product is a precision measuring instrument. We would like to point out that you should take due care when choosing the installation site and following these suggestions and instructions.
  • Page 7: General Information

    V-Flow Line General information Operating principle of the variable area flowmeter The measuring instrument works according to the float measuring principle The measuring unit consists of a tapered graduated precision glass measuring cylinder in which a ball float can move freely up and down. The medium flows through the vertically-aligned measuring cylinder from the bottom upwards.
  • Page 8: Variable Area Flow Meter Type Marking

    V-Flow Line Variable area flow meter type marking The instrument type label is attached to the inside of the left side panel in the V-100 and on the back of the mounting plate in the Q-Flow: Example: 483493-1-006 07/08 133526 Type: FLV-CSSA-SM Air, 3.6-43 ln/min, 4 bar a, 20°C...
  • Page 9: Technical Data

    V-Flow Line Technical data Technical data Q-Flow Type Q-Flow 55 Q-Flow 80 Q-Flow 140 Turndown ratio 10:1 10:1 10:1 Accuracy in % of full scale ± 5 % ± 4 % ± 2 % Measuring tube length 55 mm 80 mm 140 mm Scale length 40 mm...
  • Page 10: Technical Data V-100

    V-Flow Line Technical data V-100 Type V-100 55 V-100 80 V-100 140 Turndown ratio 10:1 10:1 10:1 Accuracy in % of full scale ± 5 % ± 4 % ± 2 % Measuring tube length 55 mm 80 mm 140 mm Scale length 40 mm 65 mm...
  • Page 11: Technical Data M-Flow

    V-Flow Line Technical data M-Flow Type M-Flow 25 M-Flow 35 M-Flow V-Stack   Straight valve  Corner valve   Valve insert/cartridge  Gas distribution system  Ccw-closed Valve turns Valve size NS 1.0 bis 3.0 NS 4 und 6.5 NS 4 und 6.5 Leak rate <1x10...
  • Page 12: M-Flow Valve Operation Options

    V-Flow Line M-Flow valve operation options The following valve operation options are available (see also the type code of each instrument): Standard-knob Standard-knob with locking ring manual knob manual knob stop ring 518-1552 basic valve right- or left-handed disc Ø10x6.4/0.5 518-1550 thread made from brass or headless screw M3x6 512-8204 headless screw M3x6 512-8204...
  • Page 13: Measuring Ranges Variable Area Flowmeters

    V-Flow Line Measuring ranges variable area flowmeters Measuring ranges direct reading scales ln/h (Air) and l/h (Water) Medium Air, ln/h Water, l/h 1.21 bara, 20°C 20°C Conditions Q-Flow 55 Q-Flow 80 Q-Flow 140 Q-Flow 55 Q-Flow 140 Type V-100 55 V-100 80 V-100 140 V-100 55...
  • Page 14: Measuring Ranges Direct Reading Scales Mln/Min And Ln/Min (Air)

    V-Flow Line Measuring ranges direct reading scales mln/min and ln/min (Air) Medium Air, mln/min Conditions 1.013 bara, 20°C Type Q-Flow 80 / V-100 80 Q-Flow 140 / V-100 140 10-120 10-100 20-240 40-220 60-640 50-540 Medium Air, ln/min Conditions 1.013 bara, 20°C Type Q-Flow 80 / V-100 80 Q-Flow 140 / V-100 140...
  • Page 15: C V -Values High-Precision Control Valves M-Flow

    V-Flow Line -values high-precision control valves M-Flow -values valves NS1.0 to NS6.5 0.01 0.001 0.0001 NS-1.0 NS-1.5 NS-2.0 0.00001 NS-2.5 NS-3.0 NS-4.0 0.000001 NS-6.5 0.0000001 spindle turns /h water at Δp of 1bar -value 1 = 1 m Manual Version Page V-Flow Line vflow_E4_7...
  • Page 16: Installation And Commissioning

    V-Flow Line Installation and commissioning General instructions Measuring instrument design In addition to the type code (see appendix) we require you to supply the following information for optimal design of the instrument:  Medium  Maximum flow rate  For design without valve: Pressure in measuring cylinder ...
  • Page 17 V-Flow Line What is the purpose of standard liters and norm liters? In contrast to liquids, gases can be compressed, so it is important to know the density of the gas used. This in turn is dependent on pressure and temperature, i.e. if two quantities of gas are to be compared with one another, then the current pressure in the pipe and the current temperature of the medium must be known.
  • Page 18: Installation Instructions

    V-Flow Line Installation instructions  Before installing the measuring instrument, the pipe must be free of particles and moisture  The variable area flowmeter must be installed in a vertical position  The direction of flow is from the bottom upwards ...
  • Page 19: Panel Mounting Q-Flow

    V-Flow Line Panel mounting Q-Flow  For mounting, the panel opening must be prepared according to the diagram  Screw the two mounting brackets onto the instrument as shown  Screw the setscrew supplied into the connector and insert the instrument into the aperture from the front ...
  • Page 20: Faceplate Mounting V-100

    V-Flow Line Faceplate mounting V-100  For mounting, the faceplate must be prepared according to the diagram  The instrument is supplied with mounted plate and inserted into the aperture from the front. Screw the plate to the control cabinet using suitable M3 screws ...
  • Page 21: Limit Indicator

    V-Flow Line Limit indicator The V-100 variable area flowmeters can be equipped with a maximum of two limit indicators. The instruments with limit indicators have the following standard design: Minimum Switch Maximum Switch Initiator position Switching status Note: In case of power failure, the contact closes irrespective of the position of the float. For initialization, the float must first pass the initiator so that it is recognized by the initiator and the float position is defined.
  • Page 22 V-Flow Line Setting the limit indicator  Remove the front glass (1)  Unscrew both clamping screws (2)  Position the limit indicator  Tighten the clamping screws  Fit the front glass For retrofit installation of a limit indicator the measuring glass must be removed.
  • Page 23 V-Flow Line Minimum spacing for limit indicators To ensure that the limit indicators do not interfere with one another a minimum space must be complied with. Installed vertically in a measuring instrument: Measurement instruments adjacent to one another: A = Minimum 16 mm B = Minimum 6mm Areas of application Limit indicators only work with steel variable area flowmeters.
  • Page 24: M-Flow Digiturn Installation

    V-Flow Line M-Flow Digiturn installation  Slide the adjusting ring 518-1563 onto the valve (do not tighten yet)  Slide the adapter sleeve 518-1575 onto the valve axis and tighten the setscrews 512-8202  Slide the rear part of the hand knob onto the adapter sleeve again. The pin on the knob must be pushed into the hole in the adjusting ring completely so that no gap remains ...
  • Page 25: Disassembly And Maintenance

    V-Flow Line Disassembly and maintenance Important instructions  Close the valves before and after the measuring instrument  Close the needle valve on the measuring instrument Caution When disassembling the measuring glass, residual liquids or gases may escape. As the float is loose inside the measuring glass, care must be taken when removing it so that the float is not lost.
  • Page 26: Disassembly Variable Area Flowmeter V-100

    V-Flow Line Disassembly variable area flowmeter V-100  Remove the screws (1) and Plexiglas (2)  Unscrew the locking screw (3) while holding the measuring glass firmly (hexagon socket, key size 5 mm)  If the screw is loosened sufficiently, then the measuring glass can be removed Refitting Refitting is carried out in the reverse sequence: ...
  • Page 27: Disassembly Digiturn M-Flow

    V-Flow Line Disassembly Digiturn M-Flow  Close the rotary knob or valve until the scale shows 00.00  Open the cover ring 633-1104 and remove from the rotary knob  Unscrew the M5x5 setscrew. The back part of the knob can be pulled off (hexagon socket, key size 2.5 mm) ...
  • Page 28: Maintenance

    V-Flow Line Maintenance When used correctly, variable area flow meters from Vögtlin Instruments AG do not require any maintenance. Soiling The following symptoms indicate soiling:  The set value can no longer be reached – the control valve is probably soiled ...
  • Page 29: Appendix

    V-Flow Line Appendix Dimensions Q-Flow Type Dimensions in mm Q-Flow 55 Q-Flow 80 Q-Flow 140 Fitting Thread depth G 1/4“ female G 1/8“ female NPT 1/4“ female G 1/4“ female for compression fittings Manual Version Page V-Flow Line vflow_E4_7 © Vögtlin Instruments AG...
  • Page 30: Dimensions V-100

    V-Flow Line Dimensions V-100 Type Dimensions in mm 120.5 89.5 61.5 V-100 55 145.5 114.5 86.5 V-100 80 205.5 174.5 146.5 V-100 140 Fitting Thread depth G 1/4“ female G 1/8“ female NPT 1/4“ female G 1/4“ female for compression fittings Manual Version Page...
  • Page 31: Dimensions M-Flow

    V-Flow Line Dimensions M-Flow M-Flow 25 straight valve Fitting Dimensions in mm Thread depth G 1/4“ female G 1/8“ female NPT 1/4“ female G 1/4“ female for compression fittings M-Flow 25 corner valve 8xM4 depth 4mm Fitting Dimensions in mm Thread depth G 1/4“...
  • Page 32 V-Flow Line M-Flow 35 straight valve mounting hole 8xM4 Gas distribution system / V-Stack 2xM4 per block Manual Version Page V-Flow Line vflow_E4_7 © Vögtlin Instruments AG...
  • Page 33: Type Code Q-Flow

    V-Flow Line Type code Q-Flow Instrument type Q-Flow Instrument size Materials (body, seals) Aluminium/brass, FKM Stainless steel, FKM Stainless steel, EPDM Customer-specific / OEM Setup Setup S** Customer-specific / OEM Valve (position) Valve at the inlet** Valve at the outlet Without valve Customer-specific / OEM G 1/4“...
  • Page 34: Type Code V-100

    V-Flow Line Type code V-100 Instrument type V-100 Instrument size Materials (body, seals) Aluminium/brass, FKM Stainless steel, FKM Stainless steel, EPDM Stainless steel, FFKM Customer-specific / OEM Setup Setup S** Setup L Setup R Setup A (stainless steel only) Setup T (stainless steel only) Setup O (stainless steel only) Customer-specific / OEM Valve (position)
  • Page 35: Type Code M-Flow

    V-Flow Line Type code M-Flow Instrument type M-Flow Instrument size V-Stack Materials (body, seals) Aluminium/brass, FKM Stainless steel, FKM Stainless steel, EPDM Stainless steel, FFKM Customer-specific / OEM Setup Straight valve** Corner valve Valve insert without body V-Stack, 1 valve/outlet V-Stack, 2 valves/outlets V-Stack, 3 valves/outlets V-Stack, 4 valves/outlets...
  • Page 36 V-Flow Line Options Locking ring (for standard-knob) Locking nut (instead of standard-knob) Digiturn (cw-closed valve only) Star grip for digiturn Digi-knob, 100 divisons (ccw-closed valve only) Hose connectors PVC, G 1/4" / 4mm Hose connectors PVC, G 1/4" / 6mm Hose connectors PVC, G 1/4"...
  • Page 37: Contamination Declaration

    V-Flow Line Contamination declaration With return of devices, please fill out the following statement completely, especially the reason for the return, the type of residue and cleaning in the case of soiling, as well as indication of hazards. Device: Type code: Serial number: Reason for return: Type of contamination...
  • Page 38 V-Flow Line Change log Date Version Replaces Author Note 01.03.2010 vflow_E4_3 vflow_E4_2 Page 21 / New Article Code 17.01.2011 vflow_E4_4 vflow_E4_3 Page 12, 21, 29, 30, 31, 32: Translation of marking Page 9, 10: Material of limit stop added 07.05.2012 vflow_E4_5 vflow_E4_4 Page 18 / Sealing Material...