T A C 7
A p p e a r a n c e
3. Push in sample drawer until it stops.
4. Push in drawer handle to engage cassette to the proper measurement location.
Push in handle
5. Start the white calibration process from the application.
6. After the white calibration is complete, open the drawer and lower the sample plate.
7. Remove the white calibration card and return it to the storage sleeve.
8. Position the color calibration card (TAC7-157) on the sample plate so that it completely
covers the sample plate.
Color calibration card on
sample plate
9. Lift the sample plate up until it stops against the measurement spot and close the cassette
clamps.
14
C a p t u r e
S y s t e m
Handle in closed
position