CALIBRATION

The software application prompts for an instrument calibration when required. Refer below for
procedure.
A calibration consists of a white tile card measurement followed by a ColorChecker target card
measurement.
The white calibration tile is dramatically affected by smudge marks, dust, and finger prints. Refer
to Appendices for calibration tile handling procedure.
Make sure to use the calibration tile supplied with the instrument for calibrating. Do not
NOTE:
substitute a calibration tile from another instrument.

Calibration Procedure

1. Open the sample drawer and lower the sample plate if required. Position the white calibration
card (TAC7-156) on the sample plate so that it completely covers the sample plate.
White calibration card on
sample plate
2. Lift the sample plate up until it stops against the measurement spot and close the cassette
clamps.
White calibration card in
measurement position
T A C 7
A p p e a r a n c e
C a p t u r e
S y s t e m
13