Attachments; Error Codes - IKA KS 4000 i control Manual

Hide thumbs Also See for KS 4000 i control:
Table of Contents

Advertisement

Water is the only coolant that may be used (with antifre-
eze, e.g. ethylene glycol).
Permissible cooling agents - inlet temperature >3°C
Observe the maximum permissible pressure of 1 bar! As
a safeguard, we recommend using a pressure limiter (e.g.
IKA C25). This is not necessary when using an IKA
KV600.
Accumulated condensate is drained out of the cooler through the
drain hose.

Attachments

The following attachments are offered as accessories
AS 4000.1 universal attachment
AS 4000.3 bowl attachment
The attachments are affixed to the shaking table using the four
clamping screws provided.
Load (standard values)
Load / kg
6
Max. speed / rpm
500
24
AS 4000.2 bracket attachment
12
15
20
400
300
120
Assembly information for AS 4000.1

Error codes

An error is indicated by an acoustic signal and an error code is dis-
played.
E.g.:
Note:
During continuous operation, the load may slip, for example, resul-
ting in the appliance becoming excessively unbalanced. To pre-
vent this, there is a motion sensor built in which reduces the
speed gradually once a critical limit has been reached.
When faults are not specified in the table, you switched the
appliance off and again on.
If faults cannot be eliminated directly, you must perform a
RESET (see "Resetting parameters to the factory settings")!
If the faults still cannot be eliminated, the appliance must be
inspected by a technical service.