Cardiac Science Powerheart G5 User Manual page 59

Automated external defibrillator
Hide thumbs Also See for Powerheart G5:
Table of Contents

Advertisement

Table A-8: CPR device (optional)
Voice prompt
Remove the green square
package from the lid of
the AED.
Tear open the green package
and remove the CPR device.
Place the CPR device on the
centre of the patient's chest,
between the nipples.
Place the heel of one hand
on the CPR device.
Press slower.
Press faster.
Press softer.
Press harder and fully release. PRESS HARDER
Powerheart® G5 Automated External Defibrillator
Text display
Line 1
Line 2
REMOVE GREEN SQUARE
PACKAGE FROM AED LID
TEAR OPEN PACKAGE
REMOVE CPR DEVICE
PLACE DEVICE ON
CENTRE OF CHEST
PLACE ONE HAND ON
CPR DEVICE
PRESS SLOWER
PRESS FASTER
PRESS SOFTER
FULLY RELEASE
Situation
Prompts at the beginning of a CPR
session. The green package contains
the CPR device.
Prompts to remove the CPR device.
Prompts the rescuer to put the CPR
device in the correct place for giving
compressions.
Prompts the rescuer to place one
hand on the CPR device.
If the rescuer is giving compressions
too quickly, prompts them to slow
the rate.
If the rescuer is giving compressions
too slowly, prompts them to quicken
the rate.
If the rescuer is giving compressions
that are too deep, prompts them to
reduce the depth.
If the rescuer is giving compressions
that are too shallow, prompts them
to use more effort and release all
pressure when moving hands up.
Prompt level
Adv Std
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
70-00569-02 G
Bas
X
X
X
X
A-11

Hide quick links:

Advertisement

Table of Contents
loading