YZ has been advised of the possibility of such damages. Information contained in this document is subject to change without notice and does not represent a commitment on the part of YZ Systems, Inc. All prices quoted are in U.S. dollars, F.O.B. Snyder, Texas. NJEX is a trademark of YZ Systems, Inc. All other product names and/or registered trademarks are the property of their respective holders.
1. Introduction Congratulations on your purchase of the DynaPak 3010RNM Series Sampler. Before installation, insure that all of the components are present. You may or may not have or- dered a sample cylinder(s) with your DynaPak System. Regardless, you will need a cylinder during installation.
3. Theory of Operation: The DynaPak 3010RNM Sampler is a liquid The Z-65/6.1 counter periodically energizes a low power solenoid valve. As in the time based sampling system which uses the pneumatically sampling mode of operation, this allows operated, positive displacement DynaPak 3000 HP pump, the Z-65/6.1 timer/controller, actuation gas to stroke the DP-3000 HP pump.
4. System Installation: CAUTION: **Incorrect operation of valves (over tightening) can result in damage to the valve components (isolation valve bonnet assembly) which might result in the valve stem being screwed out of the probe body. This of course results in product at pipeline pressure being vented continually through this port until this section of the pipeline is shut in.
4. System Installation: CAUTION: Incorrect operation of valves (over tightening) can result in damage to the valve components (isolation valve bonnet assembly) which might result in the valve stem being screwed out of the probe body. This of course results in product at pipeline pressure being vented continually through this port until this section of the pipeline is shut in.
5. Sample Vessel Installation: The free-floating piston cylinder (DuraSite) should be installed in a horizontal position on the vessel rack. Install 1/8" tubing from the sample discharge port of the manifold to the product end of the vessel. Avoid traps in this line. The vessel may be pre-charged by using bottled inert gas such as nitrogen or helium (consult the factory for proper procedure).
6. Operational Check & Leak Testing: 6.1 When all of the tubing connections have been completed, close the purge valve on the front of the sampler probe body. Open the sample probe supply valve to establish product supply pressure to the probe body. Check all connections for leaks.
7. Sampler Set-Up: Continuous time-based sampling 7.1 Calculate the sampling rate using the following chart: Number of turns sample Sample cylinder volumes open on pump pump stroke knob displacement 1000 cc 300 cc 500 cc per stroke .100 .200 Sample rate (minutes) .300...
Page 11
7. Sampler Set-Up: Continuous time-based sampling 7.3 Adjust the pump volume adjustment knob to the value used in the calculations in step 7.1. Sample pump displacement per stroke Number of turns open on the pump volume knob .1cc .2cc .4cc Factory Mutual System...
8. Sampler set-up: Continuous proportional-to-flow sampling In this mode of operation, the Z-65/6.1 controller and the PIM-100 module work together as a dividing counter to control the rate at which the pump is actuated. The desired time between pump strokes is controlled by the turbine meter pulses or another output device that will give an input pulse to the PIM-100 module.
Page 13
8. Sampler set-up: Continuous proportional-to-flow sampling To set this value turn on switch positions S2-2, S1-8, S1-7, S1-6, S1-5, S1-4, and S1-3. Turning a switch "ON" enables the divisor value. Switch divisor values are cumulative. 2300 -2048 S2-2 “ON” - 128 S1-8 “ON” - 64 S1-7 “ON”...
Page 14
8. Sampler set-up: Continuous proportional-to-flow sampling 8.2 Z-65/6.1 Set Up: Calculate the Z65/6.1 counter setting using the following chart: 1. your pump displacement (from .1 to .4cc’s) a.__________ 2. your sample cylinder volume in cc’s (300cc, 500cc, etc.) b.__________ 3. average flow rate (Metered Volume per day) c.__________ 4.
Page 15
8. Sampler set-up: Continuous proportional-to-flow sampling Number of turns open on the pump volume knob 8.3 Adjust the pump volume adjustment knob to the value used in the calculation in step 8.1. Sample pump displacement per stroke .1cc .2cc .4cc 8.4 Input from the PIM-100 is a dry contact: a.interface between the PIM-100 and the Z-65/6.1 terminal strip will be pre-wired.
9. Sampler maintenance: 9.1 Recommended preventative F. Remove the diaphragm housing from the maintenance schedule pump body by unscrewing the diaphragm Every sampling situation is unique. Below are housing and carefully pulling the plunger out of our recommendations for average conditions. the pump body.
Page 17
9. Sampler maintenance: G. Remove the internal bushings and o-rings from the pump body by inserting a nonmetallic rod (larger than 1/4", smaller than 1/2") into the top of the pump body. Gently tap to remove all bushings and o-rings out the bottom of the pump body.
Page 18
NOTE: follow the illustration to assure proper battery wire placement in the Z-65/6.1 enclo- sure. 5. Return the mode switches to their original positions. 6. Send your depleted battery to: YZ Systems Inc. 206 Lubbock Hwy. Snyder, TX 79549 USA Systems, Inc. •...
Page 19
9. Sampler maintenance: 9.4 Recommended spare parts for the DynaPak 3000RNM Series gas samplers. Part Number Description Qty. Location C4-0004 filter element see diagrams #3 and #4 D3-0115 DP-3000 HP pump seal kit see diagrams #1 and #2 D3-0116 YZ filter regulator repair kit see diagrams #3 and #4 D3-0142 Z-65/200 fuse replacement kit...
10. Troubleshooting: Timer Mode 10.1 Mechanical Operation Test: A. Set the mode switches as follows: 1. Positions 1, 2 and 3 on. B. Set the time switches to 00 to enter the diagnostic mode. This mode enables the user to increase the solenoid output rate to one pulse every two seconds.
Page 21
10. Troubleshooting: 10.4 Timer Range Setting A. the Z-65/6.1 timer has two ranges for the timer setting dials. 1. Z-65/6.1 Range Setting: a. xx minutes: set the configuration jumper to the far left position (factory setting). b. x.x minutes: set the configuration jumper to Factory Mutual System...
Page 22
10. Troubleshooting: Counter Mode 10.5 Input Pulse Test A. Set the mode switches as follows: 1. Position 1 and 3 on, 2 and 4 off. B. Set the count switches to 00 to enter the diagnostic mode. This mode enables the user to determine if the proper input pulses are being received at the count input (ter.
Diagram #8: DuraSite Portable Sample Vessel Instructions Purpose: The DuraSite Portable Sample Vessel permits the Step 3b: (Gas sampling) Connect the pre-charge port to the user to remove a liquid or gas hydrocarbon sample from a DuraSite to the pipeline for pre-charge pressure, or configured pipeline or a sampling device.
Diagram #9: About The PIM-100... Theory Of Operation: Introduction: The PIM-100 User's Guide provides all of the The PIM-100 Pulse Input Module is a low power device information you will need to work with the Pulse Input designed to interface between a DynaPak Z-65 Module.
Need help?
Do you have a question about the DynaPak 3010RNM Series and is the answer not in the manual?
Questions and answers