8.1.4 Waveform Specifications

The RESCUE SAM delivers a nominal 200J Biphasic Truncated Exponential waveform to patients with
impedances ranging from 25 to 175 ohms.
The waveform is adjusted to compensate for measured patient impedance. Nominal phase times and
energy delivered are shown in the tables below.
Patient
Impedance
(Ohms)
25
50
75
100
125
150
175
8.1.5 Patient
Analysis System
The RESCUE SAM Patient Analysis System ensures that the pad/patient impedance is within the proper
range and analyzes the patient's ECG rhythm to determine whether a shock is required. In an initial ECG
signal conditioning stage, ECG data are received and digital processing is done to remove baseline
wander and high-frequency noise.
The ECG signal are then processed by the ECG signal analysis routines. The arrhythmia detection process
determines whether or not to advise shocking the patient by examining the outputs of these analyses.
Manual
• in the case in which the patient gives vital
signs during the charging phase of the
defibrillator the recommended action is to
disconnect immediately the connector of
the
defibrillator.
I in this way the defibrillator discharges
energy internally.
Phase A,
Duration (msec)
Duration (msec)
6
8
8
10
10
12
12
RESCUE SAM User Manual vers.1.0 Jun. 2015
disposable
electrodes
Phase B,
Energy Delivered
6
200J+/-10%
6
200J+/-10%
8
200J+/-10%
8
200J+/-10%
10
200J+/-10%
10
200J+/-12%
10
200J+/-12%
from
the
(Joules)