Memory And Data Transfer; How Registration And Memory Management Work; View Events - Progetti RESCUE SAM User Manual

Bi-phasic semi-automatic external defibrillator
Table of Contents

Advertisement

5

Memory and data transfer

Rescue SAM defibrillator has an internal memory capable of recording ECG traces and events happened
during the emergency phases.
5.1

How registration and memory management work

The internal defibrillator memory is able to record up to 8 sessions, every session could be long up to 60
minutes. Once the 8 session will be complete the defibrillator will replace and rewrite automatically the
oldest session.
The data inside the device are kept also when the defibrillator is turned OFF.
Progetti s.r.l. company offers as optional the possibility to see the registration on a computer by an
interface of data download and software management.
WARNING
WARNING

5.1.1 View events

Sam Data Manager is an application software based on Windows, it is able to read the data recorded
inside the RESCUE SAM, showed and manage it on the PC. The main function for Sam Data Manager are:
The rescue staff is able by the software to rebuild a cardiac event starting from the moment when
the AED is turned ON and connected to the patient until to the RESCUE SAM where turned OFF.
It allows to medical staff to review the event in every moment.
It allows to service point and to the manufacturer a clear and detailed reconstruction of all events
and the proper use of the AED analyzing the performance of the device.
It provides to the technicians extra parameters for the detection of faults in equipment suspected
of improper operation.
WARNING
WARNING
RESCUE SAM User Manual vers.1.0 Jun. 2015
RESCUE SAM defibrillator is able to record for single
event 60 minutes. If the data collected exceed this
limit, the following ones will be not recorded.
RESCUE SAM overwrite the oldest event. Be sure to
archive the data on a PC in order to avoid to lost
past events.
Sam Data Manager 1.0 is an application software
stand-alone that could not be used when the AED is
working, it is developed only to support the analysis
post event of the data recorded on the internal
memory.
The recording of an event must be brought to the
competent medical facility in the area to allow to
examine the data.
15