PRESENTATION AND
INSTALLATION
OF DEVICES
0 766 60: 6-DIRECTION* CALL DISPLAY UNIT
Performs the main and secondary display functions.
Main display unit: used to monitor rooms in the ward from the
nurses' station.
Secondary display unit: used to transfer information from the
main display unit. 4 secondary display units max.
Indicates calls by lighting up numbered indicator lights and
generating an audible signal.
Can indicate several calls simultaneously.
Displays the call and nurse presence
and mute setting.
Calls can be transferred to 4 other display units (treatment room,
rest area, etc).
Supplied without plate.
Supplied with sheets of numbered stickers from 1 to 99.
Technical characteristics
Power supply: 24 V=
n
Consumption: min. 0.4 W - max. 2.8 W
n
Operating temperature: 5 to 40°C
n
Storage temperature: -20 to 70°C
n
Protection index: IP 40
n
Dimensions: 135 x 22.5 mm
n
Installation:
n
- On table control unit Cat. No. 0 782 14
- Flush-mounted in box depth 50 mm Cat. No. 0 800 53
- Surface-mounted in box depth 40 mm Cat. No. 0 802 86
Sound level:
- 65 dB (A) at 1 m placed on table
n
- 75 dB (A) fixed to wall 1.60 m off the floor
- 62 dB (A) at 1 m placed on table
- 72 dB (A) fixed to wall 1.60 m off the floor
* Direction: room from which the calls originate
Direction
identification*
Mute button
Connection
1
1- Management module terminals
2- Room 6 terminals
3- Room 1 terminals
4- Room 2 terminals
5- Room 3 terminals
6- Room 4 terminals
7- Room 5 terminals

PRESENTATION AND INSTALLATION OF DEVICES

Identification
labels supplied
2
0 766 60
3
4
5
6
7
MOSAIC NURSE CALL UNIT
11