Boston Scientific S-ICD System Directions For Use
Boston Scientific S-ICD System Directions For Use

Boston Scientific S-ICD System Directions For Use

Advertisement

Quick Links

S-ICD
System
TM
Patient Screening Steps
Prior to use, please review full S-ICD Directions for Use
1

Setup

• Remove hair and clean skin prior to placing surface electrode
• Start with patient laying supine
• Identify Xiphoid Process (Ask patient to "draw in" their stomach. Palpate the edge of the ribcage, moving toward the
midline, until they meet at the Xiphoid Process)
• Place electrodes starting with LA lead
• Use screening tool to determine RA lead location
2

Obtain a Clean ECG

• Establish a stable baseline
• Record 10 – 20 seconds of ECG in 2 different postures - supine and standing, or supine and sitting
• ECG recording:
- Leads: I, II and III
- Sweep speed: 25 mm/sec
- ECG gain: 5 – 20 mm/mV
• Use the largest ECG gain that does not clip the peak of the QRS complex
• To yield an acceptable signal for testing, the gain may be adjusted for each ECG lead independently
• Troubleshooting suggestions to help establish a stable baseline:
- Add ground electrode (RL)
- Ensure electrodes are new
3

Evaluate the ECG

• Select the colored profile from the Patient Screening Tool that best matches
the amplitude of the QRS complex:
- The peak of the QRS complex must fit within one of the Peak Zones as shown
- For biphasic signals, the larger peak should be used to select the profile
- ECG gains > 20 mm/mV are not permitted
• Align the left edge of the selected colored profile with the onset of the QRS complex:
- The horizontal line on the colored profile should be used as a guide for the
ECG baseline
• Evaluate each QRS complex:
- Ensure the entire QRS complex and trailing T-wave are contained within the
colored profile
- Repeat the evaluation process for all QRS complexes from all surface ECG leads
in all collected postures
4

Passing Criteria

• Evaluate ECG lead:
- ECG lead is successful if all QRS complexes pass the screening evaluation in all postures
- Morphology of QRS is consistent across both postures, supine and standing or supine and sitting
- A patient is considered suitable for implant if at least one surface ECG lead (sense vector) is acceptable for all
tested postures.
- For Questions or Concerns, please Call 1-800-CARDIAC (Fax: 1-800-474-3245)
Note: If viewing a faxed or scanned copy of an ECG, check scaling to confirm appropriate size.
- Ask patient to inhale then exhale and hold 10 sec. to manage respiratory artifact
- Add electrode gel

ECG Electrode Placement:

LL:
Placed lateral location, at the 5th intercostal
space along the mid-axillary line
LA: Placed 1cm left lateral of the xiphoid midline
RA: Placed 14 cm superior to the ECG Electrode
(14cm guide is located on screening tool)
RL: Ground: either on the right hip or leg for
best placement to avoid noise

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the S-ICD System and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Boston Scientific S-ICD System

  • Page 1 S-ICD System Patient Screening Steps ECG Electrode Placement: Placed lateral location, at the 5th intercostal space along the mid-axillary line LA: Placed 1cm left lateral of the xiphoid midline RA: Placed 14 cm superior to the ECG Electrode (14cm guide is located on screening tool) RL: Ground: either on the right hip or leg for best placement to avoid noise Prior to use, please review full S-ICD Directions for Use...
  • Page 2 Contraindications: Unipolar pacemakers are contraindicated for use with the S-ICD System. Warnings and Cautions: The S-ICD System contains sterile products for single use only. Do not resterilize. Handle the components of the SICD System with care at all times and maintain proper sterile technique.

Table of Contents