Download Print this page

Hach 9586sc Basic User Manual page 15

Oxygen scavenger analyzer
Hide thumbs Also See for 9586sc:

Advertisement

Available languages

Available languages

4. Make connections with twisted-pair shielded wire and connect the shield at the controlled
component end or at the control loop end.
• Do not connect the shield at both ends of the cable.
• Use of non-shielded cable may result in radio frequency emission or susceptibility levels higher
than allowed.
• Maximum loop resistance is 500 ohm.
5. Close the controller cover and tighten the cover screws.
6. Configure outputs in the controller.
Figure 3 Analog output connections
Connect the optional digital communication output
The manufacturer supports Modbus RS485, Modbus RS232, Profibus DPV1 and HART
communication protocols. The optional digital output module is installed in the location indicated by
item 4 in
Figure 2
on page 9. Refer to the instructions supplied with the network module for more
details.
For information about Modbus registers, go to
search Modbus registers or go to any sc200 product page.
Plumb the sample and drain lines
Connect the sample and drain lines to the fittings on the back of the panel.
1. Connect the sample line to the inlet 4/6-mm OD tube fitting.
2. Connect a drain line to the drain 6/8-mm OD tube fitting. Keep the drain line as short as possible
at atmospheric pressure.
http://www.de.hach.com
or
http://www.hach.com
English 15
and

Advertisement

loading