SECTION 2
TECHNICAL SPECIFICATIONS
2–1

Electrical Specification

2–1–1 Line Power Supply
2–1–2 Maximum Line Current
2–1–3 Permissible Line Resistance
2–1–4 Generator Output
2–1–5 Anode Current
2–1–6 Generator Power
2–1–7 Duty Cycle
D
Input voltage, single-phase:
D
Line frequency:
D
Line power: 6 kVA during exposures
D

Maximum line current:

200V: 30A
240 V: 25A
The maximum line impedance authorized are the following, depending on the mains nominal
voltage:
D
200V
––––> 0.33 Ohms
D
208V
––––> 0.36 Ohms
D
220V
––––> 0.40 Ohms
D
230V
––––> 0.44 Ohms
D
240V
––––> 0.48 Ohms
D
22 through 35 kV,
D
20 through 100 mA.
D
I max large focus = 100 mA at 30 kV,
D
I max small focus = 30 mA at 30 kV.
D
3 kW maximum.
The generator is always limited by the tube:
D
The generator can supply an average power 100 W to the tube.
D
This example represents one 30 kV/200mAs exposure each minute.
4–2
SENOGRAPHE 800T
200/208/220/230/240V ( 10%).
50 or 60 Hz ( 1 Hz).
om 2135575–100